ORIGINAL PAPER / OBSTETRICS

Ginekologia Polska

2022, vol. 93, no. 2, 121–125

Copyright © 2022 Via Medica

ISSN 0017–0011, e-ISSN 2543–6767

DOI 10.5603/GP.a2021.0085

Diagnostic accuracy of PremaQuick in detection of preterm labor in symptomatic women

Ibrahim A. Abdelazim12Osama O. Amer3Svetlana Shikanova4Bakyt Karimova4
1Department of Obstetrics, and Gynaecology, Ain Shams University, Cairo, Egypt
2Ahmadi Hospital, Kuwait Oil Company (KOC), Ahmadi, Kuwait
3Ghamra Military Hospital, Cairo, Egypt
4West Kazakhstan Marat Ospanov Medical University, Aktobe, Kazakhstan
ABSTRACT
Objectives: Failure to identify women at risk of preterm labor (PTL) leads to failure to implement standard measures. This study designed to evaluate the accuracy of PremaQuick test in detection of PTL in women presented with threatened preterm labor (TPTL).
Material and methods: One hundred and twenty-two (122) pregnant women, singleton pregnancy, < 37 weeks, admitted with TPTL included in this study, and were compared to 122 controls.
After thorough evaluation, participants were examined using sterile vaginal speculum for cervico-vaginal fluid (CVF) sampling, and PremaQuick test. The CVF sampling was followed by trans-vaginal sonographic (TVS) assessment of cervical length (CL). Participants were managed according to hospitals policy thorough their admission, and after discharge in the ante-natal clinics till delivery. After delivery, the delivery data were compared by the recorded participants` data on admission.
Results: The PremaQuick test had 95.1% sensitivity, 97.5% specificity, 97.5% positive predictive value, 95.2% negative predictive value, and 96.3% accuracy in detection of PTL. The PremaQuick had significantly higher true negative rate, specificity, positive predictive value, and overall accuracy in detection of PTL compared to CL < 25 mm (p = 0.005, 0.005, 0.01, 0.002; respectively).
Conclusions: The PremaQuick is an accurate bedside test in detection of PTL in women presented with TPTL. It had 95.1% sensitivity, 97.5% specificity, 97.5% positive predictive value, 95.2% negative predictive value, and 96.3% overall accuracy in detection of PTL. The PremaQuick had significantly higher true negative rate, specificity, positive predictive value, and overall accuracy in detection of PTL compared to CL < 25 mm.
Key words: diagnostic accuracy; PremaQuick; preterm labor
Ginekologia Polska 2022; 93, 2: 121125

Corresponding author:

Ibrahim A. Abdelazim

Department of Obstetrics, and Gynaecology, Ain Shams University, Cairo, Egypt

dr.ibrahimanwar@gmail.com

Received: 27.11.2020 Accepted: 7.03.2021 Early publication date: 29.04.2021

This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles
and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.

INTRODUCTION

Preterm labor (PTL) is an important cause of perinatal deaths, and neonatal morbidity [1–4]. PTL occurs after excessive uterine stretch (twin or triplet pregnancies), amniotic fluid infection, or chorio-decidual hemorrhage [5, 6].

The fetal fibronectin and cervical length (CL) measured by trans-vaginal sonography (TVS) are the main diagnostic tools currently used to detect PTL [7].

The CL, and fetal fibronectin have low positive predictive value, and limited accuracy to detect PTL [8, 9].

The fetal fibronectin, and insulin growth factor binding protein-1 (IGFBP-1) are amniotic fluid markers used for prediction of PTL [10].

The fetal fibronectin test has high negative predictive value (NPV) in diagnosing PTL [11, 12]. While amniotic fluid contamination, bleeding, and unprotected intercourse are associated with false fetal fibronectin results [4].

Failure to identify women at risk of PTL leads to failure to implement standard measures with subsequent increase in perinatal deaths, and neonatal morbidity. While the false positive diagnosis of PTL exposes women to unnecessarily admission, tocolysis, and corticosteroids.

The IGFBP-1 released into the cervico-vaginal fluid (CVF) during the process of chorio-decidual disruption of PTL [13–15]. The IGFBP-1 is a good negative predictor of PTL [16]. The interlukein-6 (IL-6) is a marker of sub-clinical chorioamnionitis associated with PTL [17, 18].

It is crucial to have a reliable diagnostic tool rather than the currently available tests to identify women at risk of PTL [10]. PremaQuick is a bedside test containing antibodies against three amniotic fluid markers (Native, and total IGFBP-1, and IL-6).

Objectives

This study designed to evaluate the accuracy of PremaQuick test in detection of PTL in women presented with threatened preterm labor (TPTL).

MATERIAL AND METHODS

This prospective comparative study was conducted over 15 months (June 2019 to August 2020); after ethical committees approval (approval number OB_0403_19), and registration as clinical trial (ACTRN12618001472268) [19].

One hundred and twenty-two (122) pregnant women between 2040 years` old, singleton pregnancy, < 37 weeks` gestation, admitted with TPTL were included in this study, and compared to 122 controls, after informed consent in accordance with the Declaration of Helsinki to evaluate the accuracy of PremaQuick test in detection of PTL in women presented with TPTL.

Women without medical disorders, with pregnancy, and intact fetal membranes, between 2436+6 weeks` gestation, presented with uterine contractions (34 contractions/30 minutes), each contraction lasting for30 seconds with50% cervical effacement, and < 3 cm dilated cervix were included in TPTL group.

Pregnant women without TPTL admitted under observation for fetal wellbeing assessment because of suspected intrauterine growth retardation or for blood sugar or blood pressure monitoring due to suspected diabetes or hypertensive disorders with pregnancy, were included as controls after exclusion of intrauterine growth retardation, diabetes, and hypertensive disorders with pregnancy [15].

Women37 weeks`, twin or triplet pregnancies, intrauterine growth retardation, medical disorders with pregnancy (diabetes and/or hypertension), dilated cervix3 cm, rupture of membranes (ROM), fetal anomalies or intrauterine fetal death, and/or ante-partum hemorrhage were excluded from this study.

Women delivered preterm iatrogenically due to medical disorders with pregnancy (diabetes, hypertension, or intrahepatic cholestasis) or obstetrics indications [twins, triplets, or premature rupture of fetal membranes (PROM)] [20] were also excluded from this study.

The gestational age was estimated based on the first day of LMP (last menstrual period), and confirmed by ante-natal scan done before 20 weeks` [21–23].

Participants were examined abdominally to evaluate; the fundal height, uterine contractions (frequency and duration), and fetal heart, followed by laboratory investigation according to hospitals protocol.

Participants were also examined using sterile vaginal speculum (without antiseptics or lubricant) for CVF sampling, and PremaQuick test before CL assessment, and digital examination.

The sterile swab of PremaQuick kit (Biosynex, France) was placed in the posterior vagina for 15 seconds for CVF sampling, then placed in the extraction solution provided by manufacture for 10 seconds. Three drops of the extraction solution were dispensed into the wells of test device/cassette, then the test result detected within 10 minutes, and recorded.

The presence of 3C (control) lines is important for PremaQuick test validation, and score2 means positive PremaQuick test, while score 0 or1 means negative PremaQuick test [15].

PremaQuick test is a bedside test containing antibodies against three amniotic fluid markers: Native, and total IGFBP-1, and IL-6 [15].

The CVF sampling for PremaQuick test was followed by TVS assessment of CL by sonographer blinded to participants’ clinical data using the standard guideline (to avoid potential bias) [24], and digital examination for assessment of cervical effacement, and dilatation.

Participants were managed according to hospitals policy (hospitalization, tocolysis, and corticosteroids) based on the PremaQuick test results, CL, and clinical findings.

The participants were followed in the ante-natal clinics weekly after hospital discharge till delivery. After delivery, the delivery data were compared to the recorded participants` data on admission to evaluate the accuracy of PremaQuick test in detection of PTL in women presented with TPTL.

Statistical Analysis

Statistical analysis done using Statistical Package for Social Sciences (SPSS) version 20 (Chicago, IL, USA). The Chi-square test (x2), and student (t) were used for analysis of qualitative, and quantitative variables, respectively.

The sensitivity, specificity, predictive values, and accuracy of PremaQuick test and CL in dectection of PTL were calculated and compared. The relative risk (RR) of PTL in women with positive PremaQuick test, and CL < 25 mm was also calculated. P-value < 0.05 was considered significant.

RESULTS

One hundred and twenty-two (122) pregnant women, singleton pregnancy, < 37 weeks` gestation, admitted with TPTL were compared to 122 controls in this study to evaluate the accuracy of PremaQuick test in detection of PTL in women presented with TPTL.

There was no significant difference between the TPTL group, and controls regarding the mean maternal age, and gestational age at enrollment (30.7 ± 7.1 years, and 32.4 ± 4.1 weeks vs 33.1 ± 6.3, and 35.2 ± 3.7; respectively) (p = 0.9 and 0.1; respectively).

In TPTL group, the PremaQuick test had higher true positive rate in detection of PTL compared to CL < 25 mm (95.1% (116/122) vs 71.3% (87/122); respectively), but this difference was statistically insignificant (p = 0.1).

In controls, the PremaQuick test had significantly higher true negative rate in detection of PTL compared to CL < 25 mm (97.5% (119/122) vs 56.6% (69/122); respectively) (p = 0.005) (Tab. 1).

Table 1. Accuracy of the PremaQuick test, and cervical length (CL) in detection of Preterm labor

Variables

PremaQuick

Number (%)

CL < 25 mm

Number (%)

p-value

TPTL group (122 women)

True positive (TP)

False negative (FN)

116/122 (95.1%)

6/122 (4.9%)

87/122 (71.3%)

35/122 (28.7%)

0.1

Controls (122 women)

True negative (TN)

False positive (FP)

119/122 (97.5%)

3/122 (2.5%)

69/122 (56.6%)

53/122 (43.4%)

0.005*

Sensitivity (TP ÷ TP + FN) × 100

116 ÷ (116 + 6) × 100 = (95.1%)

87 ÷ (87 + 35) × 100 = (71.3%)

0.1

Specificity (TN ÷ TN + FP) × 100

119 ÷ (119 + 3) × 100 = (97.5%)

69 ÷ (69 + 53) × 100 = (56.6%)

0.005*

Positive predictive value (PPV)

(TP ÷ TP + FP) × 100

116 ÷ (116 + 3) × 100 = (97.5%)

87 ÷ (87 + 53) × 100 = (62.1%)

0.01*

Negative predictive value (NPV)

(TN ÷ TN + FN) × 100

119 ÷ (119 + 6) × 100 = (95.2%)

69 ÷ (69 + 35) × 100 = (66.3%)

0.07

Accuracy

(TP + TN ÷ TP + TN + FP + FN) × 100

116+119 ÷ (116 + 119 + 3 + 6) × 100 = (96.3%)

87 + 69 ÷ (87 + 69 + 53 + 35) × 100 = (63.9%)

0.002*

The PremaQuick test had 95.1% sensitivity, 97.5% specificity, 97.5% positive predictive value (PPV), 95.2% negative predictive value (NPV), and 96.3% overall accuracy in detection of PTL. While the CL < 25 mm had 71.3% sensitivity, 56.6% specificity, 62.1% PPV, 66.3% NPV, and 63.9% overall accuracy in detection of PTL (Tab. 1).

The PremaQuick had significantly higher true negative rate, specificity, PPV, and overall accuracy in detection of PTL compared to CL < 25 mm (p = 0.005, 0.005, 0.01, 0.002; respectively) (Tab. 1).

The relative risk of PTL in women presented with TPTL was higher with positive PremaQuick test (RR 20.3 (95% CI:
9.29
44.36), p = 0.0001) compared to CL < 25 mm (RR 1.85 (95% CI: 1.372.49), p = 0.0001). In addition, the number of women delivered preterm after positive PremaQuick test was significantly higher than those delivered preterm after CL < 25 mm (116/119 versus 87/140; p = 0.01) (Tab. 2).

Table 2. Relative risk of PTL with positive PremaQuick test, and cervical length (CL) < 25 mm

Variables

PTL (bad outcome)

Good outcome (No PTL)

RR (95% confidence interval) p-value

PremaQuick

Positive test group (119)

Negative test group (225)

116

6

3

119

20.3 (9.2944.36) 0.0001*

CL < 25 mm

Positive group (140)

Negative group (104)

87

35

53

69

1.85 (1.372.49) 0.0001*

DISCUSSION

Failure to identify women at risk of PTL leads to failure to implement standard measures. The false positive diagnosis of PTL exposes women to unnecessarily admission, tocolysis, and corticosteroids. It is important to have a reliable diagnostic tool rather than the currently available tests to predict women at risk of PTL [10].

PremaQuick is a bedside test containing antibodies against three amniotic fluid markers (Native, and total IGFBP-1, and IL-6). Therefore, one hundred and twenty-two (122) pregnant women, < 37 weeks`, admitted with TPTL were compared to 122 controls in this study to evaluate the accuracy of PremaQuick test in detection of PTL in women presented with TPTL.

In this study, the PremaQuick test had significantly higher true negative rate in detection of PTL compared to CL < 25 mm (97.5% (119/122) vs 56.6% (69/122); respectively), (p = 0.005). It also had significantly higher specificity, PPV, and overall accuracy compared to CL < 25 mm (p = 0.005, 0.01, and 0.002; respectively) in detection of PTL.

Similarly, Abu-Faza et al. [15], found the CL < 25 mm had low specificity, and low positive predictive value in detection of PTL.

Schmitz et al. [25], also, found the CL25 had 75% senstivity, 63% specificity, and 24% PPV in detection of PTL.

Nikolova et al. [26], concluded that the PAMG-1 (placental alpha microglobulin-1) is better predictor of imminent PTL when compared with phosphorylated-IGFBP-1 (ph IGFBP-1) alone or in combination with CL.

Melchor et al. [27], found the positive predictive value of PAMG-1 was significantly higher than the phIGFBP-1 or fetal fibronectin in detection of spontaneous PTL within seven days.

The phIGFBP-1 alone as an amniotic fluid marker has limited predictive ability to detect women at risk for PTL [28]. Therefore, the PremaQuick test designed against three amniotic markers (Native, and total IGFBP-1, and IL6) to increase its accuracy in detection of PTL.

In this study, the PremaQuick test had 95.1% sensitivity, 97.5% specificity, 97.5% positive predictive value, 95.2% negative predictive value, and 96.3% overall accuracy in detection of PTL. The PremaQuick test had significantly higher specificity, positive prtedictive value, and overall accuracy in detection of PTL compared to CL < 25 mm (p = 0.005, 0.01, and 0.002; respectively).

Similarly, Asiegbu et al. [29], found the the PremaQuick test had 96.3% senstivity, 97.6% specificity, 89.7% PPV, 99.2% NPV, and 97.3% accuracy, in detection of PTL within 14 days in women with TPTL between 2836+6 weeks` gestation.

Eleje et al. [30], also found the PremaQuick test had 100.0/87.5% sensitivity, 94.1/96.9% specificity, 70.5/87.5% PPV, 100.0/96.9% NPV and 95.0/95.0% accuracy in detection of PTL within 7/14 days in women with singleton pregnancy presented with TPTL < 35 weeks`, respectively. They concluded that the PremaQuick test is an accurate test in detection of PTL in women with singleton pregnancy presented with TPTL [30].

Abu-Faza et al. [15], also found the PremaQuick test had higher specificity and positive predictive value in diagnosing PTL compared to CL.

In this study, the relative risk of PTL in women presented with TPTL was higher with positive PremaQuick test (RR 20.3 (95% CI: 9.2944.36), p = 0.0001) compared to CL < 25 mm (RR 1.85 (95% CI: 1.372.49), p = 0.0001). In addition, the number of women delivered preterm after positive PremaQuick test was significantly higher than those delivered preterm after CL < 25 mm (116/119 vs 87/140; p = 0.01).

Abu-Faza et al. [15], also found the odds, and relative risk (RR) of PTL within 714 days in symptomatic women were significantly higher for PremaQuick (12.9, and 8.4; respectively) compared to CL < 25 mm (1.4, and 1.1; respectively).

This study found the PremaQuick is an accurate bedside test in detection of PTL in symptomatic women presented with TPTL. It had 95.1% sensitivity, 97.5% specificity, 97.5% positive predictive value, 95.2% negative predictive value, and 96.3% overall accuracy in detection of PTL. Its true negative rate, specificity, PPV, and overall accuracy in detection of PTL were significantly higher than CL < 25 mm.

The current study was the first registered, prospective, comparative, multicenter study conducted to evaluate the accuracy of PremaQuick test in detection of PTL in symptomatic women presented with TPTL.

Women refused to give consent and participate, and shipping of the PremaQuick kits were the limitations faced during this study.

The accuracy of PremaQuick test in detection of PTL should be compared with other amniotic fluid markers as PAMG-1 (AmniSure test) or IGFBP-1 (Actim-PROM test) in future studies.

CONCLUSIONS

The PremaQuick is an accurate bedside test in detection of PTL in women presented with TPTL. It had 95.1% sensitivity, 97.5% specificity, 97.5% positive predictive value, 95.2% negative predictive value, and 96.3% overall accuracy in detection of PTL. The PremaQuick had significantly higher true negative rate, specificity, positive predictive value, and overall accuracy in detection of PTL compared to CL < 25 mm.

Acknowledgments

To Dr. Thierry Paper, Deputy General Manager of Biosynex, France (Dr. Thierry supplied the PremaQuick kits freely and he was not included in any part of the study), and to women who gave consent and participate in the study.

Conflict of interest

No conflict of interests related to this study.

REFERENCES

  1. Alexander GR, Kogan M, Bader D, et al. US birth weight/gestational age-specific neonatal mortality: 19951997 rates for whites, hispanics, and blacks. Pediatrics. 2003; 111(1): e61e66, doi: 10.1542/peds.111.1.e61, indexed in Pubmed: 12509596.
  2. Martin JA, Kung HC, Mathews TJ, et al. Annual summary of vital statistics: 2006. Pediatrics. 2008; 121(4): 788801, doi: 10.1542/peds.2007-3753, indexed in Pubmed: 18381544.
  3. Goffinet F, Kayem G. Diagnosis and prognosis of preterm labor: physical examination and ultrasonography [article in French]. J Gynecol Obstet Biol Reprod (Paris). 2002; 31(Suppl 7): 5S225S34, indexed in Pubmed: 12454623.
  4. Goldenberg RL, Iams JD, Mercer BM, et al. What we have learned about the predictors of preterm birth. Seminars in Perinatology. 2003; 27(3): 185193, doi: 10.1016/s0146-0005(03)00017-x.
  5. Romero R, Mazor M, Munoz H, et al. The preterm labor syndrome. Ann N Y Acad Sci. 1994; 734: 414429, doi: 10.1111/j.1749-6632.1994.tb21771.x, indexed in Pubmed: 7978942.
  6. Lockwood CJ, Kuczynski E. Risk stratification and pathological mechanisms in preterm delivery. Paediatr Perinat Epidemiol. 2001; 15(Suppl 2): 7889, doi: 10.1046/j.1365-3016.2001.00010.x, indexed in Pubmed: 11520402.
  7. Gomez R, Romero R, Medina L, et al. Cervicovaginal fibronectin improves the prediction of preterm delivery based on sonographic cervical length in patients with preterm uterine contractions and intact membranes. Am J Obstet Gynecol. 2005; 192(2): 350359, doi: 10.1016/j.ajog.2004.09.034, indexed in Pubmed: 15695971.
  8. Sotiriadis A, Papatheodorou S, Kavvadias A, et al. Transvaginal cervical length measurement for prediction of preterm birth in women with threatened preterm labor: a meta-analysis. Ultrasound Obstet Gynecol. 2010; 35(1): 5464, doi: 10.1002/uog.7457, indexed in Pubmed: 20014326.
  9. Son M, Miller ES. Predicting preterm birth: cervical length and fetal fibronectin. Semin Perinatol. 2017; 41(8): 445451, doi: 10.1053/j.semperi.2017.08.002, indexed in Pubmed: 28935263.
  10. Georgiou HM, Di Quinzio MKW, Permezel M, et al. Predicting preterm labour: current status and future prospects. Dis Markers. 2015; 2015: 435014, doi: 10.1155/2015/435014, indexed in Pubmed: 26160993.
  11. Compan C, Rossi A, Piquier-Perret G, et al. Prediction of spontaneous preterm birth in symptomatic patients: a review [article in French]. Journal de Gynécologie Obstétrique et Biologie de la Reproduction. 2015; 44(8): 740751, doi: 10.1016/j.jgyn.2015.06.008.
  12. Leitich H, Egarter C, Kaider A, et al. Cervicovaginal fetal fibronectin as a marker for preterm delivery: a meta-analysis. Am J Obstet Gynecol. 1999; 180(5): 11691176, doi: 10.1016/s0002-9378(99)70612-5.
  13. Lembet A, Eroglu D, Ergin T, et al. New rapid bed-side test to predict preterm delivery: phosphorylated insulin-like growth factor binding protein-1 in cervical secretions. Acta Obstet Gynecol Scand. 2002; 81(8): 706712, doi: 10.1034/j.1600-0412.2002.810804.x, indexed in Pubmed: 12174153.
  14. Kekki M, Kurki T, Paavonen J, et al. Insulin-like growth factor binding protein-1 in cervix as a marker of infectious complications in pregnant women with bacterial vaginosis. The Lancet. 1999; 353(9163): 1494, doi: 10.1016/s0140-6736(99)00400-6.
  15. Abu-Faza M, Abdelazim I, Svetlana S, et al. Diagnostic accuracy of PremaQuick versus actim partus in prediction of preterm labor in symptomatic women within 14 days. Open Journal of Obstetrics and Gynecology. 2018; 8: 741755, doi: 10.4236/ojog.2018.88078.
  16. Woodworth A, Moore J, G’Sell C, et al. Diagnostic accuracy of cervicovaginal interleukin-6 and interleukin-6: albumin ratio as markers of preterm delivery. Clinical Chemistry. 2007; 53(8): 15341540, doi: 10.1373/clinchem.2007.084798.
  17. Taylor BD, Holzman CB, Fichorova RN, et al. Inflammation biomarkers in vaginal fluid and preterm delivery. Hum Reprod. 2013; 28(4): 942952, doi: 10.1093/humrep/det019, indexed in Pubmed: 23416276.
  18. Abdelazim I. Relation between interleukin-6 in the cervicovaginal fluid and subclinical chorioamnionitis in patients with preterm premature rupture of membranes. Asian Pacific Journal of Reproduction. 2013; 2(1): 3841, doi: 10.1016/s2305-0500(13)60113-x.
  19. Diagnostic accuracy of PremaQuick in detection of preterm labor in symptomatic women presented with threatened preterm labor. http://www.ANZCTR.org.au/ACTRN12618001472268.aspx (2020-03-25).
  20. Wong AE, Grobman WA. Medically indicated-iatrogenic prematurity. Clin Perinatol. 2011; 38(3): 423439, doi: 10.1016/j.clp.2011.06.002, indexed in Pubmed: 21890017.
  21. Abdelazim IA, Abdelrazak KM, Al-Kadi M, et al. Fetal fibronectin (Quick Check fFN test) versus placental alpha microglobulin-1 (AmniSure test) for detection of premature rupture of fetal membranes. Arch Gynecol Obstet. 2014; 290(3): 457464, doi: 10.1007/s00404-014-3225-5, indexed in Pubmed: 24715211.
  22. Abdelazim IA. Insulin-like growth factor binding protein-1 (Actim PROM test) for detection of premature rupture of fetal membranes. J Obstet Gynaecol Res. 2014; 40(4): 961967, doi: 10.1111/jog.12296, indexed in Pubmed: 24612210.
  23. Abdelazim IA, Makhlouf HH. Placental alpha microglobulin-1 (AmniSure test) versus insulin-like growth factor binding protein-1 (Actim PROM test) for detection of premature rupture of fetal membranes. J Obstet Gynaecol Res. 2013; 39(6): 11291136, doi: 10.1111/jog.12045, indexed in Pubmed: 23718844.
  24. Lim K, Butt K, Crane J, et al. Ultrasonographic cervical length assessment in predicting preterm birth in singleton pregnancies. Journal of Obstetrics and Gynaecology Canada. 2011; 33(5): 486499, doi: 10.1016/s1701-2163(16)34884-8.
  25. Schmitz T, Maillard F, Bessard-Bacquaert S, et al. Selective use of fetal fibronectin detection after cervical length measurement to predict spontaneous preterm delivery in women with preterm labor. Am J Obstet Gynecol. 2006; 194(1): 138143, doi: 10.1016/j.ajog.2005.05.074, indexed in Pubmed: 16389023.
  26. Nikolova T, Uotila J, Nikolova N, et al. Prediction of spontaneous preterm delivery in women presenting with premature labor: a comparison of placenta alpha microglobulin-1, phosphorylated insulin-like growth factor binding protein-1, and cervical length. Am J Obstet Gynecol. 2018; 219(6): 610.e1610.e9, doi: 10.1016/j.ajog.2018.09.016, indexed in Pubmed: 30240653.
  27. Melchor JC, Khalil A, Wing D, et al. Prediction of preterm delivery in symptomatic women using PAMG-1, fetal fibronectin and phIGFBP-1 tests: systematic review and meta-analysis. Ultrasound Obstet Gynecol. 2018; 52(4): 442451, doi: 10.1002/uog.19119, indexed in Pubmed: 29920825.
  28. Conde-Agudelo A, Romero R. Cervical phosphorylated insulin-like growth factor binding protein-1 test for the prediction of preterm birth: a systematic review and metaanalysis. Am J Obstet Gynecol. 2016; 214(1): 5773, doi: 10.1016/j.ajog.2015.06.060, indexed in Pubmed: 26149828.
  29. Asiegbu AC, Eleje GU, Ibeneme EM, et al. Combined insulin-like growth factor binding protein-1/interleukin-6 (Premaquick) versus fetal fibronectin for predicting preterm delivery among women with preterm contractions. Int J Gynaecol Obstet. 2020; 149(2): 171177, doi: 10.1002/ijgo.13128, indexed in Pubmed: 32090329.
  30. Eleje GU, Ezugwu EC, Eke AC, et al. Accuracy of a combined insulin-like growth factor-binding protein-1/interleukin-6 test (Premaquick) in predicting delivery in women with threatened preterm labor. J Perinat Med. 2017; 45(8): 915924, doi: 10.1515/jpm-2016-0339, indexed in Pubmed: 28236632.

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., ul. Świętokrzyska 73, 80–180 Gdańsk
tel.:+48 58 320 94 94, faks:+48 58 320 94 60, e-mail:  viamedica@viamedica.pl