Vasoconstrictive responses of the cephalic vein during first-time cardiac implantable electronic device placement

R. Steckiewicz1, E. B. Świętoń1, M. Bogdańska1, P. Stolarz1
·
Pubmed: 29345717
·
Folia Morphol 2018;77(3):464-470.
Affiliations
  1. Department of Cardiology, Medical University of Warsaw, Poland, Banacha 1a, Warsaw, Poland

open access

Vol 77, No 3 (2018)
ORIGINAL ARTICLES
Submitted: 2017-09-20
Accepted: 2017-12-14
Published online: 2018-01-08

Abstract

Background: During cardiac implantable electronic device (CIED) implantation procedures cardiac leads have been mostly introduced transvenously. The associated injury to the selected vessel and adjacent tissues may induce reflex vasoconstriction. The aim of the study was to assess the incidence of cephalic vein (CV) vasoconstriction during first-time CIED implantation.
Materials and methods: Of the 146 evaluated first-time CIED implantation procedures conducted in our centre in 2016, we selected those during which CV vasoconstriction was recorded. We focused on the stage of the procedure involving CV cutdown and/or axillary vein (AV)/subclavian vein (SV) puncture for lead insertion. Only cases documented via venography were considered.
Results: Vasoconstriction was observed in 11 patients (5 females and 6 males, mean age 59.0 ± 21.2 years). The presence of this phenomenon affected the stage of CIED implantation involving cardiac lead insertion to the venous system, in severe cases, requiring a change of approach from CV cutdown to AV/SV puncture. The extent of vasoconstriction front propagation was limited to the nearest valves. Histological examinations of collected CV samples revealed an altered spatial arrangement of myocytes in the tunica media at the level of leaflet attachment.
Conclusions: Cephalic vein vasoconstriction is a rare phenomenon associated with accessing the venous system during first-time CIED implantation. The propagation of CV constriction was limited by the location of the nearest valves.

Abstract

Background: During cardiac implantable electronic device (CIED) implantation procedures cardiac leads have been mostly introduced transvenously. The associated injury to the selected vessel and adjacent tissues may induce reflex vasoconstriction. The aim of the study was to assess the incidence of cephalic vein (CV) vasoconstriction during first-time CIED implantation.
Materials and methods: Of the 146 evaluated first-time CIED implantation procedures conducted in our centre in 2016, we selected those during which CV vasoconstriction was recorded. We focused on the stage of the procedure involving CV cutdown and/or axillary vein (AV)/subclavian vein (SV) puncture for lead insertion. Only cases documented via venography were considered.
Results: Vasoconstriction was observed in 11 patients (5 females and 6 males, mean age 59.0 ± 21.2 years). The presence of this phenomenon affected the stage of CIED implantation involving cardiac lead insertion to the venous system, in severe cases, requiring a change of approach from CV cutdown to AV/SV puncture. The extent of vasoconstriction front propagation was limited to the nearest valves. Histological examinations of collected CV samples revealed an altered spatial arrangement of myocytes in the tunica media at the level of leaflet attachment.
Conclusions: Cephalic vein vasoconstriction is a rare phenomenon associated with accessing the venous system during first-time CIED implantation. The propagation of CV constriction was limited by the location of the nearest valves.

Get Citation

Keywords

cephalic vein; venous spasm; tunica media; venotomy; puncture; lead implantation; cardiac implantable electronic device

Supp./Additional Files (1)
Figure 5.
View
332KB
About this article
Title

Vasoconstrictive responses of the cephalic vein during first-time cardiac implantable electronic device placement

Journal

Folia Morphologica

Issue

Vol 77, No 3 (2018)

Article type

Original article

Pages

464-470

Published online

2018-01-08

Page views

1676

Article views/downloads

806

DOI

10.5603/FM.a2018.0001

Pubmed

29345717

Bibliographic record

Folia Morphol 2018;77(3):464-470.

Keywords

cephalic vein
venous spasm
tunica media
venotomy
puncture
lead implantation
cardiac implantable electronic device

Authors

R. Steckiewicz
E. B. Świętoń
M. Bogdańska
P. Stolarz

References (13)
  1. Bongiorni MG, Proclemer A, Dobreanu D, et al. Preferred tools and techniques for implantation of cardiac electronic devices in Europe: results of the European Heart Rhythm Association survey. Europace. 2013; 15(11): 1664–1668.
  2. Cooper RM, Krishnan U, Pyatt JR. Central venous spasm during pacemaker insertion. Heart. 2010; 96(18): 1484.
  3. Duan Xu, Ling F, Shen Y, et al. Venous spasm during contrast-guided axillary vein puncture for pacemaker or defibrillator lead implantation. Europace. 2012; 14(7): 1008–1011.
  4. Jacobs DM, Fink AS, Miller RP, et al. Anatomical and morphological evaluation of pacemaker lead compression. Pacing Clin Electrophysiol. 1993; 16(3 Pt 1): 434–444.
  5. Kodani T, Mine T, Kishima H, et al. Spontaneous subclavian venous occlusion before electronic device implantation. Asian Cardiovasc Thorac Ann. 2015; 23(5): 530–534.
  6. Kolettis TM, Lysitsas DN, Apostolidis D, et al. Improved 'cut-down' technique for transvenous pacemaker lead implantation. Europace. 2010; 12(9): 1282–1285.
  7. Loukas M, Myers CS, Wartmann ChT, et al. The clinical anatomy of the cephalic vein in the deltopectoral triangle. Folia Morphol. 2008; 67(1): 72–77.
  8. Neri R, Cesario AS, Baragli D, et al. Permanent pacing lead insertion through the cephalic vein using an hydrophilic guidewire. Pacing Clin Electrophysiol. 2003; 26(12): 2313–2314.
  9. Newton DJ, McLeod GA, Khan F, et al. Mechanisms influencing the vasoactive effects of lidocaine in human skin. Anaesthesia. 2007; 62(2): 146–150.
  10. Oginosawa Y, Abe H, Nakashima Y. Prevalence of venous anatomic variants and occlusion among patients undergoing implantation of transvenous leads. Pacing Clin Electrophysiol. 2005; 28(5): 425–428.
  11. Steckiewicz R, Górko D, Świętoń EB, et al. Axillary vein spasm during cardiac implantable electronic device implantation. Folia Morphol. 2016; 75(4): 543–549.
  12. Tse HF, Lau CP, Leung SK. A cephalic vein cutdown and venography technique to facilitate pacemaker and defibrillator lead implantation. Pacing Clin Electrophysiol. 2001; 24(4 Pt 1): 469–473.
  13. Wali MA, Eid RA, Dewan M, et al. Pre-existing histopathological changes in the cephalic vein of renal failure patients before arterio-venous fistula (AVF) construction. Ann Thorac Cardiovasc Surg. 2006; 12(5): 341–348.

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Grupa Via Medica, Świętokrzyska 73, 80–180 Gdańsk, Poland

tel.: +48 58 320 94 94, faks: +48 58 320 94 60, e-mail: viamedica@viamedica.pl