Axillary vein spasm during cardiac implantable electronic device implantation

R. Steckiewicz, D. Górko, E. B. Świętoń, G. Szparecki, P. Stolarz
·
Pubmed: 27830883
·
Folia Morphol 2016;75(4):543-549.

open access

Vol 75, No 4 (2016)
ORIGINAL ARTICLES
Submitted: 2016-02-29
Accepted: 2016-04-05
Published online: 2016-05-10

Abstract

Background: The technique of axillary vein (AV) or subclavian vein (SV) puncture has become an important alternative to cephalic vein (CV) cutdown as an approach allowing cardiac lead introduction into the venous system during cardiac implantable electronic device (CIED) implantation procedures. Irrespective of the technique used, the injury associated with lead insertion may induce a reflex venous spasm that can even cause total venous obstruction. In order to assess the incidence of AV spasm during AV puncture, we analysed a total of 735 (382 in females and 353 in males; mean age 75 ± 11 years) de novo CIED implantation procedures involving transvenous lead insertion conducted between January 2014 and December 2015.

Materials and methods: In 337 patients the leads were introduced via AV puncture only, in 66 patients AV puncture was used in combination with CV cutdown, together yielding a total of 403 procedures (55% of all de novo CIED implantation procedures; mean patient age 72 ± 14 years), out of which we observed 12 cases (mean patient age 57 ± 25 years) of AV spasm (3%).

Results: We evaluated only the procedures with unambiguous fluoroscopy images recorded during AV puncture: complete blockage of contrast medium flow through the AV, with preserved flow through the CV or collateral vessels, followed by eventually resumed flow of contrast via the AV. The contrast-enhanced movements of AV walls showed the spasm propagating both proximally and distally along the vessel, while the subsequent vessel wall relaxation occurred along the entire spasm-affected venous segment simultaneously.

Conclusions: An AV spasm induced by AV puncture during CIED implantation is a rare phenomenon; however, if severe, it may significantly affect the course of the procedure  

Abstract

Background: The technique of axillary vein (AV) or subclavian vein (SV) puncture has become an important alternative to cephalic vein (CV) cutdown as an approach allowing cardiac lead introduction into the venous system during cardiac implantable electronic device (CIED) implantation procedures. Irrespective of the technique used, the injury associated with lead insertion may induce a reflex venous spasm that can even cause total venous obstruction. In order to assess the incidence of AV spasm during AV puncture, we analysed a total of 735 (382 in females and 353 in males; mean age 75 ± 11 years) de novo CIED implantation procedures involving transvenous lead insertion conducted between January 2014 and December 2015.

Materials and methods: In 337 patients the leads were introduced via AV puncture only, in 66 patients AV puncture was used in combination with CV cutdown, together yielding a total of 403 procedures (55% of all de novo CIED implantation procedures; mean patient age 72 ± 14 years), out of which we observed 12 cases (mean patient age 57 ± 25 years) of AV spasm (3%).

Results: We evaluated only the procedures with unambiguous fluoroscopy images recorded during AV puncture: complete blockage of contrast medium flow through the AV, with preserved flow through the CV or collateral vessels, followed by eventually resumed flow of contrast via the AV. The contrast-enhanced movements of AV walls showed the spasm propagating both proximally and distally along the vessel, while the subsequent vessel wall relaxation occurred along the entire spasm-affected venous segment simultaneously.

Conclusions: An AV spasm induced by AV puncture during CIED implantation is a rare phenomenon; however, if severe, it may significantly affect the course of the procedure  

Get Citation

Keywords

venous spasm, axillary vein, venipuncture, venography, cardiac implantable electronic device

About this article
Title

Axillary vein spasm during cardiac implantable electronic device implantation

Journal

Folia Morphologica

Issue

Vol 75, No 4 (2016)

Article type

Original article

Pages

543-549

Published online

2016-05-10

Page views

1333

Article views/downloads

1164

DOI

10.5603/FM.a2016.0027

Pubmed

27830883

Bibliographic record

Folia Morphol 2016;75(4):543-549.

Keywords

venous spasm
axillary vein
venipuncture
venography
cardiac implantable electronic device

Authors

R. Steckiewicz
D. Górko
E. B. Świętoń
G. Szparecki
P. Stolarz

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Grupa Via Medica, Świętokrzyska 73, 80–180 Gdańsk, Poland

tel.: +48 58 320 94 94, faks: +48 58 320 94 60, e-mail: viamedica@viamedica.pl