Vol 2, No 4 (2001): Practical Diabetology
Other materials agreed with the Editors
Published online: 2001-11-19
Detection of hypoglycemia with the GlucoWatch biographer
Diabetologia Praktyczna 2001;2(4):307-314.
Abstract
INTRODUCTION. Hypoglycemia is a common acute
complication of diabetes therapy. The GlucoWatch
biographer provides frequent and automatic glucose
measurements with an adjustable low-glucose
alarm. We have analyzed the performance of the
biographer low-glucose alarm relative to hypoglycemia
as defined by blood glucose £ 3.9 mmol/l.
MATERIAL AND METHODS. The analysis was based on 1,091 biographer uses from four clinical trials, which generated 14,487 paired (biographer and blood glucose) readings.
RESULTS. The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose £ 3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose > 3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer.
CONCLUSIONS. The frequent and automatic nature of the biographer readings allows more effective detection of hypoglycemia than that achieved with current medical practice.
MATERIAL AND METHODS. The analysis was based on 1,091 biographer uses from four clinical trials, which generated 14,487 paired (biographer and blood glucose) readings.
RESULTS. The results show that as the low-glucose alert level of the biographer is increased, the number of true positive alerts (alarm sounds and blood glucose £ 3.9 mmol/l) and false positive alerts (alarm sounds but blood glucose > 3.9 mmol/l) increased. When analyzed as a function of varying low-glucose alert levels, the results show receiver operator characteristic curves consistent with a highly useful diagnostic tool. Setting the alert level from 1.1 to 1.7 mmol/l above the level of concern is likely to optimize the trade-off between true positives and false positives for each user. When the same blood glucose data are analyzed for typical monitoring practices (two or four measurements per day), the results show that fewer hypoglycemic events are detected than those detected with the biographer.
CONCLUSIONS. The frequent and automatic nature of the biographer readings allows more effective detection of hypoglycemia than that achieved with current medical practice.
Keywords: hypoglycemiaglycemia measurementsbiographerionophoresis