Comparison of neoatherosclerosis and clinical outcomes between bioabsorbable versus durable polymer drug-eluting stent: Verification by optical coherence tomography analysis

Jae Young Cho*1Hyungdon Kook*2Javoxir Anvarov3Najmiddin Makhkamov3Sang-A Cho4Cheol Woong Yu5
1Division of Cardiology, Department of Internal Medicine, Regional Cardiocerebrovascular Center, Wonkwang University Hospital, Iksan, Korea
2Division of Cardiology, Department of Internal Medicine, College of Medicine, Hanyang University, Seoul, Korea
3Department of Endovascular Surgery and Cardiac Arrhythmias, Republican Specialized Center of Surgery named after academician V. Vakhidov, Tashkent, Uzbekistan
4Health Insurance Review and Assessment Service, Wonju, Korea
5Division of Cardiology, Department of Internal Medicine, Korea University Anam Hospital, Seoul, Korea

Address for correspondence: Cheol Woong Yu, MD, PhD, Department of Cardiology, Korea University Anam Hospital, 73 Inchon-ro, Seongbuk-gu, Seoul, Korea 02841, tel: +82-2-920-5445, fax: +82-2-927-1418, e-mail: ycw717@naver.com

Received: 7.12.2020 Accepted: 11.10.2021 Early publication date: 19.04.2022

*Jae Young Cho and Hyungdon Kook are co-first authors.

This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.

Abstract
Background: Neoatherosclerosis after drug-eluting stent (DES) implantation is known to be related with increased risk of late restenosis and stent thrombosis. Neoatherosclerosis and relevant clinical outcomes between bioabsorbable polymer DES (BP-DES) and second-generation durable polymer DES (DP-DES) were evaluated by optical coherence tomography (OCT) analysis.
Methods: A total of 311 patients (319 lesions) undergoing OCT analysis after DES implantation were enrolled and divided into two groups according to stent type (BP-DES [150 patients, 153 lesions] and DP-DES [161 patients, 166 lesions]). Follow-up OCT analysis was performed at least 9 months after index stent implantation. Neoatherosclerosis was defined as presence of thin-cap fibroatheroma, calcified plaque, and lipid plaque. Primary endpoint was the incidence of neoatherosclerosis, and the secondary endpoints were the occurrence of major adverse cardiac events (MACE), defined as a composite of death, myocardial infarction, target lesion revascularization, or stent thrombosis and to find independent predictors of neoatherosclerosis.
Results: The incidence of neoatherosclerosis was lower in the BP-DES group than the DP-DES group (5.2% vs. 14.5%, p = 0.008), which was driven by lipid plaque. However, the incidence of MACE did not show statistical difference between the two groups in median 4-year follow-up (3.3% vs. 7.8%, hazard ratio 1.964, 95% confidence interval 0.688–5.611, p = 0.207). Less use of angiotensin converting enzyme inhibitors/angiotensin II receptor blockade and higher degree of neointimal hyperplasia remained independent predictors of neoatherosclerosis on Cox regression analysis.
Conclusions: Patients undergoing BP-DES implantation had lower incidence of neoatherosclerosis than DP-DES, which did not reach statistically better clinical outcomes. (Cardiol J 2023; 30, 6: 911–920)
Key words: neointima, coronary restenosis, drug-eluting stents, tomography, optical coherence

Introduction

Introduction of drug-eluting stent (DES) have markedly reduced in-stent restenosis and repeat revascularizations. Yet, stent failure related to stent thrombosis or restenosis was a remnant of major concern, which in turn may be related to fatal clinical events [1, 2]. Development of neoatherosclerosis within neointimal tissues is one of the main mechanisms of stent failure, and the incidence of neoatherosclerosis in first-generation DES is reported from 36% to 67% [3–5] of cases. Second-generation durable polymer DES (DP-DES) have maintained the low restenosis rates of first-generation DES with reduced rates of stent failure. Recently, however, very late stent thrombosis and neoatherosclerosis, resulting in adverse clinical outcomes, have been observed with second-generation DP-DES [1, 6]. One mechanism of late stent failure has been attributed to the delayed endothelial healing secondary to a hypersensitivity reaction to the durable polymer [7, 8]. Development of bioabsorbable polymer DES (BP-DES) was one of the attempts to overcome this problem. Although most of the polymer degradation process is complete within 6–9 months, several studies have shown that polymers requiring active bioresorption are associated with higher rates of inflammation than durable polymer [9, 10]. There is no clear evidence and it remains controversial concerning the benefit of BP-DES on neoatherosclerosis development and clinical events after overcoming the inflammatory process during bioresorption [11–16]. The aim of the present study was to compare the incidence of neoatherosclerosis, relevant clinical outcomes after stent implantation between BP-DES and second-generation DP-DES by using optical coherence tomography (OCT).

Methods

Study patients

The Korea University Anam Hospital OCT Registry is a single-center registry of patients undergoing OCT imaging of the coronary arteries (ClinicalTrials.gov identifier NCT02966262). From OCT registry database, 656 lesions in 630 patients were retrospectively screened from March 2011 through April 2015. The inclusion criteria of the present study were as follows: 1) Patients who underwent percutaneous coronary intervention (PCI) with BP-DES or second-generation DP-DES; 2) Patients who underwent OCT follow-up, showing the mean neointimal thickness was > 100 µm. The reasons for follow-up angiography and OCT were evidence of myocardial ischemia or symptoms of coronary artery disease or planned follow-up angiography for other stented lesions. Patients were excluded if first-generation DES or bare-metal stent was implanted; clinical event occurred before OCT analysis; time interval of OCT after PCI was less than 9 months or more than 36 months. The reason that cases of time interval of OCT after PCI less than 9 months were excluded was to examine the effect of polymer-free stent state, which needs a 6–9 month absorption period of bioabsorbable polymer coating [7]. Patients were then allocated into the BP-DES group or DP-DES group regarding stent polymer type. Stents used in the current study were biolimus-eluting stents (Nobori®, Terumo Corporation, Japan; Biomatrix®, Biosensors International, Singapore) in BP-DES and everolimus-eluting stents (Xience®, Abbott Vascular, USA) in DP-DES.

Demographic data, prescribed drugs, laboratory data, and clinical presentation, were collected based on the electronic chart review and were compared between the two groups. Primary endpoint was the incidence of neoatherosclerosis based on OCT analysis, and the secondary endpoints were the occurrence of major adverse cardiac events (MACE; a composite of death, myocardial infarction, target lesion revascularization, or stent thrombosis) and to find independent predictors of neoatherosclerosis. Stent thrombosis was defined according to the recommendations of the Academic Research Consortium [17]. Information on clinical outcome was collected by a retrospective review of the chart. This study was approved by the Korea University Hospital Institute Review Board (IRB No. 2016AN0095), and informed consent was waived due to the retrospective study design. This study also complied with the Declaration of Helsinki.

Angiographic analysis

Coronary angiograms were analyzed using a computer-based telecardiology system, version 2.02 (Medcon Inc., Tel Aviv, Israel) by three radiologic technicians who were blinded to the purpose of the study. The reference diameter, minimal luminal diameter, percentage of stenosis, and lesion length were evaluated from diastolic frames using guided catheter magnification calibration in a single, matched view with a computerized quantitative analyzer using a caliper. The average diameter of normal segments proximal and distal to the treated lesion was used as the reference diameter.

OCT acquisition

Optical coherence tomography examination and analysis were performed immediately after stent implantation (LightLab Imaging Inc., Ilumien Offline review workstation, Ver E.4.1, MA, USA). Using a 0.014” guide wire, an OCT imaging catheter (C7 DragonflyTM, LightLab Imaging Inc., MA, USA) was advanced into the distal end of the DES implantation site. The entire length of the stent was imaged with an automatic pullback device moving at 15 mm/s. The whole stent could be clearly visualized on each OCT image; in-segment cross-sectional views were also obtained.

OCT analysis

All baseline OCT images were reviewed by an independent observer who was blinded to the clinical presentation, lesion, and procedural characteristics. The analysis encompassed the intra-stent segment, defined by the first and the last cross-sections with a visible strut, and the adjacent vessel segments 5 mm proximal and distal to the stent, defined as edge segments. The stent and lumen areas were traced, and minimum, maximum, and mean neointimal thickness were semiautomatically determined. Plaque characteristics and restenotic tissue patterns were analyzed if the mean neointimal thickness was > 100 µm on ≥ 3 consecutive cross-sectional frames at a 1 mm interval.

Lipid plaque was defined as a diffusely bordered signal-poor region with rapid signal attenuation. In lipid plaque, lipid arc was measured at every 1-mm interval throughout the entire length of each lesion and the values were averaged. Lipid length was also measured on longitudinal view. Lipid index was defined as the averaged lipid arc multiplied by lipid length [18]. Calcified plaque was defined as a clearly delineated signal-poor region with low backscatter. Thin-cap fibroatheroma was defined as neointima with a fibrous cap thickness at the thinnest part ≤ 65 µm and an angle of lipid-laden neointima ≥ 180° [4, 19]. The stent was considered to have neoatherosclerosis when lipid plaque, calcified plaque or thin-cap fibroatheroma were present [20, 21]. Microvessels were defined as well-delineated low backscattering structures < 200 μm in diameter showing a trajectory within the vessel [22]. Evagination was identified as outward bulges in the luminal contour between struts, with the depth of the bulge exceeding the actual strut thickness [23, 24]. For restenotic tissue pattern analysis, parameters were defined as follows: 1) homogeneous neointima, a uniform signal-rich band without focal variation or attenuation; 2) heterogeneous neointima, focally changing optical properties and various backscattering patterns; 3) layered neointima, layers with different optical properties, namely an adluminal high scattering layer and abluminal low scattering layer [22]. Representative OCT images of the neoatherosclerosis and in-stent restenosis pattern are visualized in Figure 1.

Figure 1. Representative optical coherence tomography (OCT) findings; A. Lipid plaque; B. Calcified plaque; C. Thin cap fibroatheroma; D. Microvessel; E. Evagination; F. Homogeneous pattern; G. Heterogeneous pattern; H. Layered pattern.
Statistical analysis

Data are expressed as mean ± standard deviation for continuous variables, whereas data for categorical variables are expressed as number and percentage of patients. The χ2 test was used to compare categorical variables. Continuous variables were compared using a t test or the Mann-Whitney test.

Selected variables were tested for univariate logistic regression associated with neoatherosclerosis; if the p-value < 0.2, they were simultaneously entered into a multivariate logistic regression model to identify independent predictors of outcome and to calculate their adjusted odds ratios with an associated 95% confidence interval (CI). The logistic regression model included the following variables, which were considered to be related with neoatherosclerosis: age, male sex, hypertension, diabetes mellitus, chronic kidney disease, smoking status, low density lipid cholesterol, high sensitive C-reactive protein, statin use, renin–angiotensin system (RAS) blocker use, stent length and stent polymer type.

Clinical event rates were estimated using the Kaplan-Meier survival analysis at 4 years, and hazard ratios (HRs) were generated using the Cox regression analysis. Because patients may have experienced more than 1 MACE, each patient was assessed until the occurrence of his or her first event and only once during analysis.

The Cox multivariate analysis to compare MACE incidence between the two study groups was performed using same covariates mentioned above as well as clinical diagnosis, incidence of neoatherosclerosis, restenotic tissue pattern, lipid index and evagination. SPSS version 24.0 (IBM SPSS Statistics, IBM Corporation, Armonk, New York) was used for all analyses. A p-value < 0.05 was considered statistically significant.

Results

The study protocol is diagrammed in Figure 2. Overall, 150 patients (153 lesions) in BP-DES group and 161 patients (166 lesions) in DP-DES group were analyzed. The baseline characteristics of the patients included in this study are presented in Table 1. Baseline characteristics of the BP-DES group were not statistically significantly dif- ferent from those of the DP-DES group except when using antiplatelet agents and an onset diagnosis. The onset diagnosis of stable angina was more frequent in the BP-DES group than in the DP-DES group (61.4% vs. 47.6%, p = 0.044). Table 2 shows the quantitative coronary angiogra-phy results. Shorter stent length was used (20.3 ± 5.6 mm vs. 22.2 ± 7.6 mm, p = 0.011) in the BP-DES group. Otherwise, other angiographic data were well matched in the study.

Figure 2. Study flow chart; PCI percutaneous coronary intervention; DES drug-eluting stent; BP-DES bioabsorbable polymer drug-eluting stent; DP-DES durable polymer drug-eluting stent; MACE major adverse cardiac events; OCT optical coherence tomography.
Table 1. Baseline patient characteristics.

Variable

BP-DES (n = 150)

DP-DES (n = 161)

P

Age [year]

60.2 ± 10

60.5 ± 9.6

0.726

Male sex

108 (72.0%)

113 (70.2%)

0.711

Body mass index [kg/m2]

25.3 ± 2.7

24.8 ± 3.0

0.206

Smoking:

0.187

Previous

11 (7.3%)

19 (11.8%)

Current

33 (22.0%)

44 (27.3%)

Comorbidity:

Hypertension

86 (57.3%)

92 (57.1%)

1.000

Diabetes mellitus

31 (20.7%)

47 (29.3%)

0.122

Chronic kidney disease

6 (4.0%)

12 (7.5%)

0.232

Laboratory data:

White blood cell count [×103/μL]

7.3 ± 2.3

8.2 ± 3.1

0.008

Creatinine [mg/dL]

0.96 ± 0.19

0.98 ± 0.43

0.578

Total cholesterol [mg/dL]

171.2 ± 37.6

169.7 ± 40.6

0.772

Triglyceride [mg/dL]

144 ± 77.8

124.4 ± 65.3

0.038

HDL-C [mg/dL]

44.6 ± 10.5

45 ± 11.4

0.753

LDL-C [mg/dL]

99.9 ± 34.6

93.4 ± 26.5

0.115

Peak CK-MB [ng/mL]

27.8 ± 79.5

40.4 ± 84.4

0.305

hs-CRP [mg/dL]

5.558 ± 21.433

10.398 ± 33.544

0.248

ESR [mm/h]

9.8 ± 7.8

11 ± 10.3

0.387

HbA1c [%]

6.8 ± 1.2

6.5 ± 1.0

0.248

Drug:

Acetylsalicylic acid

136 (90.7%)

158 (97.5%)

0.013

P2Y12 inhibitor

148 (98.7%)

151 (93.2%)

0.021

Cilostazol

8 (5.3%)

8 (4.9%)

1.000

Statin

144 (96.0%)

153 (94.4%)

0.603

RAS blocker

71 (47.3%)

89 (54.9%)

0.212

Beta-blocker

90 (60.0%)

107 (66.0%)

0.292

Calcium channel blocker

66 (44.0%)

64 (39.5%)

0.424

Clinical presentation:

0.044

Stable angina

92 (61.3%)

77 (47.8%)

Unstable angina

35 (23.3%)

50 (31.1%)

NSTEMI

13 (8.7%)

25 (15.5%)

STEMI

10 (6.7%)

9 (5.6%)

Table 2. Angiographic and procedural data.

Variable

BP-DES (153 lesions)

DP-DES (166 lesions)

P

Vessel:

0.940

LAD

96 (62.7%)

108 (65.1%)

LCX

23 (15.0%)

25 (15.1%)

RCA

30 (19.6%)

30 (18.1%)

Left main

4 (2.6%)

3 (1.8%)

Mean stent diameter [mm]

3.00 ± 0.44

3.03 ± 0.44

0.563

Mean stent length [mm]

20.3 ± 5.6

22.2 ± 7.6

0.011

Quantitative coronary analysis:

Baseline:

Proximal RD [mm]

3.0 ± 0.3

2.9 ± 0.5

0.907

Distal RD [mm]

2.3 ± 1.16

2.4 ± 1.1

0.554

MLD [mm]

0.8 ± 0.6

0.9 ± 0.6

0.791

Diameter stenosis [%]

68.6 ± 27.3

69.0 ± 21.8

0.941

Post-procedure:

Proximal RD [mm]

3.1 ± 0.4

3.2 ± 0.5

0.433

Distal RD [mm]

2.9 ± 0.4

3.1 ± 0.5

0.227

MLD [mm]

2.8 ± 0.4

2.9 ± 0.5

0.377

Diameter stenosis [%]

5.7 ± 3.7

5.8 ± 4.8

0.867

OCT findings

Optical coherence tomography findings at follow-up are shown in Table 3. Although time from stent implantation to follow-up OCT was similar (12.1 ± 2.3 months vs. 12.6 ± 2.4 months, p = 0.062), the prevalence of neoatherosclerosis was lower in the BP-DES group compared with the DP-DES group (5.2% vs. 14.5%, p = 0.008), which was driven by lipid plaque (5.2% vs. 13.9%, p = 0.013). However, lipid index was numerically higher in the BP-DES group but showed no statistical difference between the two groups (1385.5 ± 2324.5 vs. 636.8 ± 984.6, p = 0.403). In quantitative analysis, mean neointimal area as well as mean neointimal thickness were similar among each group. In other hand, incidence (62.5% vs. 30.3%, p < 0.001), frame count (24.0 ± 16.8 vs. 12.2 ± 8.3, p < 0.001) and frame rate from total frame (28.1 ± 19.0% vs. 13.3 ± 8.8%, p < 0.001) of stent evagination had all presented significantly higher in the BP-DES group. The intra-observer k coefficient for OCT findings was 0.91, and the interobserver κ coefficient was 0.90.

Table 3. Comparison of OCT findings.

Variable

BP-DES (153 lesions)

DP-DES (166 lesions)

P

Index PCI to OCT duration [month]

12.1 ± 2.3

12.6 ± 2.4

0.062

OCT analysis:

Mean stent area [mm2]

6.39 ± 2.64

6.88 ± 2.59

0.227

Minimal luminal area [mm2]

4.59 ± 2.03

4.99 ± 2.29

0.249

Mean neointimal area [mm2]

1.80 ± 0.96

1.89 ± 1.05

0.554

Mean neointimal thickness [mm]

0.22 ± 0.09

0.24 ± 0.13

0.374

Neoatherosclerosis:

8 (5.2%)

24 (14.5%)

0.008

Microvessel

2 (1.7%)

3 (2.3%)

0.377

Thin cap fibroatheroma

0 (0%)

2 (1.2%)

0.499

Calcified plaque

1 (0.7%)

0 (0%)

0.478

Lipid plaque

8 (5.2%)

23 (13.9%)

0.013

Macrophage

5 (4.2%)

10 (7.4%)

0.300

Lipid plaque:

Lipid plaque length [mm]

8.03 ± 9.97

4.54 ± 4.22

0.366

Lipid plaque arc [°]

109.6 ± 63.3

108.8 ± 67.7

0.978

Lipid index

1385.5 ± 2324.5

636.8 ± 984.6

0.403

ISR pattern:

< 0.001

Homogeneous

35 (22.9%)

79 (47.6%)

Heterogeneous

16 (10.5%)

27 (16.3%)

Layered

11 (7.2%)

14 (8.4%)

Evagination:

95 (62.5%)

50 (30.3%)

< 0.001

Evagination frame

24.0 ± 16.8

12.2 ± 8.3

< 0.001

Evagination rate [%]

28.1 ± 19.0

13.3 ± 8.8

< 0.001

Predictors of neoatherosclerosis

Bioabsorbable polymer DES failed to predict neoatherosclerosis in the univariate model; only mean neointimal thickness showed statistical association. The multivariate model identified less use of RAS blocker and higher degree of neointimal hyperplasia as independent predictors of neoatherosclerosis (Table 4).

Table 4. Predictors of neoatherosclerosis.

Variable

Univariate analysis

Multivariate analysis

OR

95% CI

P

OR

95% CI

P

Male sex

0.574

0.217

1.517

0.263

Age

0.991

0.956

1.027

0.614

Diabetes mellitus

1.186

0.461

3.055

0.724

Hypertension

1.203

0.512

2.826

0.671

CKD

1.020

0.239

4.344

0.979

Smoking:

Current

2.849

1.092

7.433

0.032

Previous

1.171

0.453

3.031

0.745

UA

1.241

0.345

4.471

0.741

MI

1.949

0.429

8.862

0.388

hs-CRP

1.003

0.983

1.024

0.762

Triglyceride

1.001

0.995

1.007

0.654

LDL-C

1.004

0.990

1.018

0.604

Statin

1.552

0.210

11.459

0.666

RAS blocker

0.511

0.234

1.118

0.093

0.388

0.175

0.859

0.020

BP-DES

2.177

0.917

5.170

0.078

Stent diameter

1.019

0.437

2.376

0.966

Stent length

1.031

0.977

1.088

0.271

Mean neointimal thickness

9.178

1.349

62.454

0.023

Mean neointimal area

1.377

1.049

1.809

0.021

1.318

1.012

1.717

0.040

Clinical outcomes

Although the follow-up period was shorter in the BP-DES group (months after index PCI: 44.1 ± 13.2 vs. 51.2 ± 17.9, p < 0.001; months after follow-up OCT: 31.4 ± 13.2 vs. 38.1 ± 17.6, p < 0.001), the incidence of MACE was very low and there was no significant difference between the two groups (3.3% vs. 7.8%, HR 1.964, 95% CI 0.688–5.611, p = 0.207) (Fig. 3, Table 5). Also, the incidence of each MACE component did not differ significantly between the two groups.

Figure 3. Kaplan-Meier curve regarding cumulative survival free of major adverse cardiac events (MACE); BP-DES bioabsorbable polymer drug-eluting stent; DP-DES durable polymer drug-eluting stent.
Table 5. Clinical outcomes.

Variable

BP-DES (n = 150)

DP-DES (n = 161)

P

MACE

5 (3.3%)

13 (7.8%)

0.207

Non-fatal MI

0 (0%)

2 (1.2%)

0.531

All cause death

0 (0%)

4 (2.4%)

0.124

TLR

4 (2.6%)

8 (4.8%)

0.432

TVR

5 (3.3%)

9 (5.4%)

0.477

Non-TLR, TVR

12 (7.8%)

14 (8.4%)

0.284

Stent thrombosis

0 (0%)

3 (1.8%)

0.435

MACE after index PCI [month]

44.1 ± 13.1

51.2 ± 17.9

< 0.001

MACE after OCT [month]

31.4 ± 13.2

38.1 ± 17.6

< 0.001

Discussion

This study showed that incidence of neoatherosclerosis was lower in the BP-DES group compared to the DP-DES group, which was driven by lipid plaque. However, the difference of neoatherosclerosis incidence among stent polymer type did not transform into incidence difference of clinical events over the follow-up (median 49 months). Independent predictors of neoatherosclerosis were less use of RAS blocker and higher degree of neointimal hyperplasia.

Development of atherosclerosis in neointima, so-called neoatherosclerosis is a significant risk for late stent failure such as very late stent thrombosis or stent restenosis [4]. Pathological studies of first-generation DES have reported that the stent polymer might be the key of neoatherosclerosis acceleration by inducing inflammation in the vessel wall [25]. In advance, BP-DES was developed to overcome the risk of vessel wall damage by the durable polymer. However, benefit of BP-DES is still controversial. BP-DES showed superiority of neoatherosclerosis incidence compared to first-generation DES [14]. Meta-analysis has shown BP-DES were superior to second-generation DES in late lumen loss and late stent thrombosis [11]. However, other reports presented the incidence of neither neoatherosclerosis or clinical events were improved or were even worse with BP-DES compared to second-generation DES [12, 13, 15, 16]. The present study revealed lower incidence of neoatherosclerosis, mainly lipid plaque, in BP-DES. However, quantitative measurement of lipid plaque by lipid index was similar between the two groups. Lipid index is known to be related with plaque vulnerability [18, 26]. Since clinical event by neoatherosclerosis is mostly related with stent thrombosis, it can suggest the severity of neoatherosclerosis in terms of plaque vulnerability and that the existence of neoatherosclerosis only plays more of a role in future events by plaque rupture.

An animal-based pathological study compared BP-DES with platinum-chromium scaffold (Synergy®, Boston Scientific, USA), BP-DES with stainless scaffold (Nobori®) and second-generation DP-DES (Resolute Integrity, Medtronic, USA) concerning about neoatherosclerosis [27]. Inflammation score and neoatherosclerosis was lowest for BP-DES with platinum-chromium scaffold, followed by BP-DES with stainless scaffold and was the highest for second-generation DP-DES. Although both BP-DES were superior compared to second-generation DP-DES, there was also significant difference among two different BP-DES. Since BP-DES with platinum-chromium scaffold has strength with more biocompatible stent scaffold and a shorter degradation period than BP-DESs which were used in the current study, current findings about BP-DES may have a weak point in terms of biocompatibility. Another issue is about bioabsorbable polymer itself. Several animal-based pathologic studies have shown that bioabsorbable polymer are associated with higher rates of inflammation than durable polymer [9, 10, 28]. In the present study, evagination was less frequently observed in second-generation DP-DES than BP-DES. Evagination is known to be associated with late-acquired positive vessel remodeling and increased risk of late stent thrombosis [23, 24, 29, 30]. The polymer absorbing process may cause hypersensitivity and vessel wall inflammation leading to evagination. This finding could possibly explain the attenuated benefit of BP-DES.

Although predictors of neoatherosclerosis differed in previous reports, less use of RAS blocker and higher degree of neointimal thickness were independently correlated with neoatherosclerosis in several reports including the present study [21, 31, 32]. It is well known that activation of renin–angiotensin–aldosterone system promotes vascular inflammation and remodeling and therefore RAS blocker inhibits atherosclerosis as reduced plaque burden in atherosclerotic vessels [33]. In addition to the anti-inflammatory effect, RAS blocker has a protective effect of neointimal growth based on the findings of angiotensin II and angiotensin converting enzyme facilitates neointimal formation [5, 34]. Moreover, homogeneous and layered pattern of neointima which is linked with neointimal stabilization were more likely to be present in the use of RAS blocker in a single center OCT study [35]. With these synergistic effects, such mechanisms can possibly explain protective effect of RAS blocker on neoatherosclerosis.

Limitations of the study

This study had several limitations. This was a non-randomized retrospective study based on a relatively limited sample size and modest follow-up period, raising the possibility of selection bias and was therefore underpowered to clarify the incidence as well as clinical outcome of neoatherosclerosis. A few baseline demographics were mismatched between the BP-DES and DP-DES group, which may introduce a confounder in the statistical analysis and lead to a biased result. Also, as mentioned above, more biocompatible BP-DES was introduced yet available during study enrollment. Therefore, results of the present study should not be simplified to express no benefit of BP-DES as a class effect. The reasons for follow-up coronary angiography were mainly evidence of inducible myocardial ischemia or symptoms of coronary artery disease. Therefore, the incidence of neoatherosclerosis was derived from a biased population and caution is needed in extrapolating these results. Finally, all measurements were performed manually, meaning that a certain degree of manual error was present. Automatic plaque analysis on OCT imaging by artificial intelligence may be useful in reducing subjectivity in image interpretation and facilitate OCT quantification of neoatherosclerosis [36]. Larger studies with a longer follow-up duration with more biocompatible BP-DES are needed to confirm the relationships between neoatherosclerosis and stent polymer.

Conclusions

At 1 year after stent implantation, neoatherosclerosis was less frequently observed in BP-DES compared with DP-DES. However, this difference did not reach significant clinical outcome difference after up to 4-year follow-up. Future studies with a larger number of participants are warranted to confirm the relationship between stent polymer type and clinical outcomes.

Acknowledgments

This study was supported by Wonkwang University in 2021.

Conflict of interest: None declared

References

  1. Räber L, Magro M, Stefanini GG, et al. Very late coronary stent thrombosis of a newer-generation everolimus-eluting stent compared with early-generation drug-eluting stents: a prospective cohort study. Circulation. 2012; 125(9): 1110–1121, doi: 10.1161/CIRCULATIONAHA.111.058560, indexed in Pubmed: 22302840.
  2. Yamaji K, Kimura T, Morimoto T, et al. Very long-term (15 to 20 years) clinical and angiographic outcome after coronary bare metal stent implantation. Circ Cardiovasc Interv. 2010; 3(5): 468–475, doi: 10.1161/CIRCINTERVENTIONS.110.958249, indexed in Pubmed: 20823392.
  3. Hong SJ, Lee SY, Hong MKi. Clinical implication of optical coherence tomography-based neoatherosclerosis. J Korean Med Sci. 2017; 32(7): 1056–1061, doi: 10.3346/jkms.2017.32.7.1056, indexed in Pubmed: 28581259.
  4. Kang SJ, Mintz GS, Akasaka T, et al. Optical coherence tomographic analysis of in-stent neoatherosclerosis after drug-eluting stent implantation. Circulation. 2011; 123(25): 2954–2963, doi: 10.1161/CIRCULATIONAHA.110.988436, indexed in Pubmed: 21646494.
  5. Lee DY, Won KJ, Lee KPa, et al. Angiotensin II facilitates neointimal formation by increasing vascular smooth muscle cell migration: Involvement of APE/Ref-1-mediated overexpression of sphingosine-1-phosphate receptor 1. Toxicol Appl Pharmacol. 2018; 347: 45–53, doi: 10.1016/j.taap.2018.03.032, indexed in Pubmed: 29609002.
  6. Dores H, Raposo L, Campante Teles R, et al. Stent thrombosis with second- versus first-generation drug-eluting stents in real-world percutaneous coronary intervention: analysis of 3806 consecutive procedures from a large-volume single-center prospective registry. J Invasive Cardiol. 2013; 25(7): 330–336, indexed in Pubmed: 23813060.
  7. Grube E, Buellesfeld L. BioMatrix Biolimus A9-eluting coronary stent: a next-generation drug-eluting stent for coronary artery disease. Expert Rev Med Devices. 2006; 3(6): 731–741, doi: 10.1586/17434440.3.6.731, indexed in Pubmed: 17280537.
  8. Smits PC, Hofma S, Togni M, et al. Abluminal biodegradable polymer biolimus-eluting stent versus durable polymer everolimus-eluting stent (COMPARE II): a randomised, controlled, non-inferiority trial. Lancet. 2013; 381(9867): 651–660, doi: 10.1016/S0140-6736(12)61852-2, indexed in Pubmed: 23374650.
  9. De Scheerder IK, Wilczek KL, Verbeken EV, et al. Biocompatibility of biodegradable and nonbiodegradable polymer-coated stents implanted in porcine peripheral arteries. Cardiovasc Intervent Radiol. 1995; 18(4): 227–232, doi: 10.1007/BF00239417, indexed in Pubmed: 8581902.
  10. van der Giessen WJ, Lincoff AM, Schwartz RS, et al. Marked inflammatory sequelae to implantation of biodegradable and nonbiodegradable polymers in porcine coronary arteries. Circulation. 1996; 94(7): 1690–1697, doi: 10.1161/01.cir.94.7.1690, indexed in Pubmed: 8840862.
  11. Lupi A, Gabrio Secco G, Rognoni A, et al. Meta-analysis of bioabsorbable versus durable polymer drug-eluting stents in 20,005 patients with coronary artery disease: an update. Catheter Cardiovasc Interv. 2014; 83(6): E193–E206, doi: 10.1002/ccd.25416, indexed in Pubmed: 24478247.
  12. Katayama Y, Kubo T, Akasaka T, et al. Two-year vascular responses to drug-eluting stents with biodegradable polymer versus durable polymer: An optical coherence tomography sub-study of the NEXT. J Cardiol. 2017; 70(6): 530–536, doi: 10.1016/j.jjcc.2017.04.005, indexed in Pubmed: 28536015.
  13. El-Hayek G, Bangalore S, Casso Dominguez A, et al. Meta-Analysis of randomized clinical trials comparing biodegradable polymer drug-eluting stent to second-generation durable polymer drug-eluting stents. JACC Cardiovasc Interv. 2017; 10(5): 462–473, doi: 10.1016/j.jcin.2016.12.002, indexed in Pubmed: 28279314.
  14. Kuramitsu S, Sonoda S, Yokoi H, et al. Long-term coronary arterial response to biodegradable polymer biolimus-eluting stents in comparison with durable polymer sirolimus-eluting stents and bare-metal stents: five-year follow-up optical coherence tomo­- graphy study. Atherosclerosis. 2014; 237(1): 23–29, doi: 10.1016/j. atherosclerosis.2014.08.031, indexed in Pubmed: 25190308.
  15. Lee Stephen WL, Tam Frankie CC, Lam Simon CC, et al. The OCT-ORION Study. Circulation: Cardiovascular Interventions. 2018; 11(4): e006034, doi: 10.1161/circinterventions.117.006034.
  16. Palmerini T, Biondi-Zoccai G, Della Riva D, et al. Clinical outcomes with bioabsorbable polymer- versus durable polymer-based drug-eluting and bare-metal stents: evidence from a comprehensive network meta-analysis. J Am Coll Cardiol. 2014; 63(4): 299–307, doi: 10.1016/j.jacc.2013.09.061, indexed in Pubmed: 24211507.
  17. Cutlip DE, Windecker S, Mehran R, et al. Clinical end points in coronary stent trials: a case for standardized definitions. Circulation. 2007; 115(17): 2344–2351, doi: 10.1161/CIRCULATIONAHA.106.685313, indexed in Pubmed: 17470709.
  18. Kato K, Yonetsu T, Kim SJ, et al. Nonculprit plaques in patients with acute coronary syndromes have more vulnerable features compared with those with non-acute coronary syndromes: a 3-vessel optical coherence tomography study. Circ Cardiovasc Imaging. 2012; 5(4): 433–440, doi: 10.1161/CIRCIMAGING.112.973701, indexed in Pubmed: 22679059.
  19. Takano M, Yamamoto M, Inami S, et al. Appearance of lipid-laden intima and neovascularization after implantation of bare-metal stents extended late-phase observation by intracoronary optical coherence tomography. J Am Coll Cardiol. 2009; 55(1): 26–32, doi: 10.1016/j.jacc.2009.08.032, indexed in Pubmed: 20117359.
  20. Lee SY, Shin DH, Mintz GS, et al. Optical coherence tomography-based evaluation of in-stent neoatherosclerosis in lesions with more than 50% neointimal cross-sectional area stenosis. EuroIntervention. 2013; 9(8): 945–951, doi: 10.4244/EIJV9I8A158, indexed in Pubmed: 24384291.
  21. Yonetsu T, Kato K, Kim SJ, et al. Predictors for neoatherosclerosis: a retrospective observational study from the optical coherence tomography registry. Circ Cardiovasc Imaging. 2012; 5(5): 660–666, doi: 10.1161/CIRCIMAGING.112.976167, indexed in Pubmed: 22798521.
  22. Gonzalo N, Serruys PW, Okamura T, et al. Optical coherence tomography patterns of stent restenosis. Am Heart J. 2009; 158(2): 284–293, doi: 10.1016/j.ahj.2009.06.004, indexed in Pubmed: 19619707.
  23. Radu MD, Pfenniger A, Räber L, et al. Flow disturbances in stent-related coronary evaginations: a computational fluid-dynamic simulation study. EuroIntervention. 2014; 10(1): 113–123, doi: 10.4244/EIJV10I1A18, indexed in Pubmed: 24213249.
  24. Radu MD, Räber L, Heo J, et al. Natural history of optical coherence tomography-detected non-flow-limiting edge dissections following drug-eluting stent implantation. EuroIntervention. 2014; 9(9): 1085–1094, doi: 10.4244/EIJV9I9A183, indexed in Pubmed: 24064426.
  25. Kitabata H, Kubo T, Komukai K, et al. Effect of strut thickness on neointimal atherosclerotic change over an extended follow-up period (≥ 4 years) after bare-metal stent implantation: intracoronary optical coherence tomography examination. Am Heart J. 2012; 163(4): 608–616, doi: 10.1016/j.ahj.2012.01.007, indexed in Pubmed: 22520527.
  26. Vergallo R, Ren X, Yonetsu T, et al. Pancoronary plaque vulnerability in patients with acute coronary syndrome and ruptured culprit plaque: a 3-vessel optical coherence tomography study. Am Heart J. 2014; 167(1): 59–67, doi: 10.1016/j.ahj.2013.10.011, indexed in Pubmed: 24332143.
  27. Nakazawa G, Torii S, Ijichi T, et al. Comparison of vascular responses following new-generation biodegradable and durable polymer-based drug-eluting stent implantation in an atherosclerotic rabbit iliac artery model. J Am Heart Assoc. 2016; 5(10), doi: 10.1161/JAHA.116.003803, indexed in Pubmed: 27792651.
  28. Potnis PA, Tesfamariam B, Wood SC. Induction of nicotinamide-adenine dinucleotide phosphate oxidase and apoptosis by biodegradable polymers in macrophages: implications for stents. J Cardiovasc Pharmacol. 2011; 57(6): 712–720, doi: 10.1097/FJC.0b013e31821a4f1e, indexed in Pubmed: 21436724.
  29. Guagliumi G, Sirbu V, Musumeci G, et al. Examination of the in vivo mechanisms of late drug-eluting stent thrombosis: findings from optical coherence tomography and intravascular ultrasound imaging. JACC Cardiovasc Interv. 2012; 5(1): 12–20, doi: 10.1016/j.jcin.2011.09.018, indexed in Pubmed: 22230145.
  30. Räber L, Zanchin T, Baumgartner S, et al. Long-term vascular healing in response to sirolimus- and paclitaxel-eluting stents: an optical coherence tomography study. JACC Cardiovasc Interv. 2012; 5(9): 946–957, doi: 10.1016/j.jcin.2012.05.012, indexed in Pubmed: 22995882.
  31. Song L, Mintz GS, Yin D, et al. Neoatherosclerosis assessed with optical coherence tomography in restenotic bare metal and first- and second-generation drug-eluting stents. Int J Cardiovasc Imaging. 2017; 33(8): 1115–1124, doi: 10.1007/s10554-017-1106-2, indexed in Pubmed: 28281026.
  32. Vergallo R, Yonetsu T, Uemura S, et al. Correlation between degree of neointimal hyperplasia and incidence and characteristics of neoatherosclerosis as assessed by optical coherence tomography. Am J Cardiol. 2013; 112(9): 1315–1321, doi: 10.1016/j.amjcard.2013.05.076, indexed in Pubmed: 23891431.
  33. Peters S. Inhibition of atherosclerosis by angiotensin II type 1 receptor antagonists. Am J Cardiovasc Drugs. 2013; 13(4): 221–224, doi: 10.1007/s40256-013-0017-4, indexed in Pubmed: 23609531.
  34. Rakugi H, Kim DK, Krieger JE, et al. Induction of angiotensin converting enzyme in the neointima after vascular injury. Possible role in restenosis. J Clin Invest. 1994; 93(1): 339–346, doi: 10.1172/JCI116965, indexed in Pubmed: 8282805.
  35. Cho JY, Hong SJ, Lim DS. Effects of angiotensin receptor blockers on neointimal characteristics in angina patients requiring stent implantation: optical coherence tomography analysis. BMC Cardiovasc Disord. 2017; 17(1): 278, doi: 10.1186/s12872-017-0709-9, indexed in Pubmed: 29141596.
  36. Chu M, Jia H, Gutiérrez-Chico JL, et al. Artificial intelligence and optical coherence tomography for the automatic characterisation of human atherosclerotic plaques. EuroIntervention. 2021; 17(1): 41–50, doi: 10.4244/EIJ-D-20-01355, indexed in Pubmed: 33528359.