Clinical outcomes of cryoballoon ablation for pulmonary vein isolation: Impact of intraprocedural heart rhythm

Bruno Reissmann12, Christian-H. Heeger23, Karena Opitz2, Michael Schlüter4, Peter Wohlmuth4, Laura Rottner12, Thomas Fink23, Jin-Hong Gerds-Li5, Shibu Mathew2, Christine Lemes2, Tilman Maurer2, Feifan Ouyang2, Karl-Heinz Kuck2, Andreas Rillig12, Doreen Schöppenthau5, Andreas Metzner12
Pubmed: 33140384
Cardiol J 2022;29(5):807-814.

Abstract

Background: The current study sought to assess the impact of the intraprocedural heart rhythm (sinus rhythm [SR] vs. atrial fibrillation [AF]) on acute procedural characteristics, durability of pulmonary vein isolation (PVI) and long-term clinical outcomes of cryoballoon (CB) ablation.

Methods: A total of 195 patients with symptomatic paroxysmal (n = 136) or persistent AF (n = 59) underwent CB-based PVI. Ablation procedures were either performed in SR (SR group; n = 147) or during AF (AF group; n = 48). Persistent AF was more frequent in the AF group than in the SR group (62% vs. 20%). All other patient baseline characteristics did not differ between the two groups.

Results: The nadir temperature during the CB applications was significantly lower in the AF group than in patients in the SR group (–49 [interquartile range, –44; –54]°C vs. –47 [–42; –52]°C, p = 0.002). Median procedure and fluoroscopy times as well as the rate of real-time recordings were not different between the two groups. Repeat ablation for the treatment of atrial arrhythmia recurrence was performed in 60 patients (SR: 44 [30%] patients; AF: 16 [33%] patients), with a trend towards a lower rate of pulmonary vein reconnections in the AF group (p = 0.07). There was no difference in 3-year arrhythmia-free survival (p = 0.8).

Conclusions: Cryoballoon-based PVI during AF results in lower nadir balloon temperatures and a trend towards a higher durability of PVI as compared to procedures performed in SR. The rate of real-time PVI recordings was not affected by the intraprocedural heart rhythm.

clinicAL CARDIOLOGY

Original article

Cardiology Journal

2022, Vol. 29, No. 5, 807–814

DOI: 10.5603/CJ.a2020.0147

Copyright © 2022 Via Medica

ISSN 1897–5593

eISSN 1898018X

Clinical outcomes of cryoballoon ablation for pulmonary vein isolation: Impact of intraprocedural heart rhythm

Bruno Reissmann12Christian-H. Heeger23Karena Opitz2Michael Schlüter4Peter Wohlmuth4Laura Rottner12Thomas Fink23Jin-Hong Gerds-Li5Shibu Mathew2Christine Lemes2Tilman Maurer2Feifan Ouyang2Karl-Heinz Kuck2Andreas Rillig12Doreen Schöppenthau*5Andreas Metzner*12
1Department of Cardiology, University Heart Center Hamburg, Germany
2Department of Cardiology, Asklepios Klinik St. Georg, Hamburg, Germany
3University Heart Center Lübeck, Germany
4Asklepios Proresearch, Hamburg, Germany
5Department of Internal Medicine and Cardiology, German Heart Institute, Berlin, Germany
170921.png

Address for correspondence: Bruno Reissmann, MD, Department of Cardiology, University Heart Center Hamburg, Martinistr. 52, 20246 Hamburg, Germany, tel: +49 (0)40 7410-58320, fax: +49 (0)40 7410-55862, e-mail: b.reissmann@uke.de

Received: 28.09.2020 Accepted: 22.10.2020 Early publication date: 30.10.2020

*Shared last authorship

This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.

Abstract

Background: The current study sought to assess the impact of the intraprocedural heart rhythm (sinus rhythm [SR] vs. atrial fibrillation [AF]) on acute procedural characteristics, durability of pulmonary vein isolation (PVI) and long-term clinical outcomes of cryoballoon (CB) ablation.

Methods: A total of 195 patients with symptomatic paroxysmal (n = 136) or persistent AF (n = 59) underwent CB-based PVI. Ablation procedures were either performed in SR (SR group; n = 147) or during AF (AF group; n = 48). Persistent AF was more frequent in the AF group than in the SR group (62% vs. 20%). All other patient baseline characteristics did not differ between the two groups.

Results: The nadir temperature during the CB applications was significantly lower in the AF group than in patients in the SR group (–49 [interquartile range, –44; –54]°C vs. –47 [–42; –52]°C, p = 0.002). Median procedure and fluoroscopy times as well as the rate of real-time recordings were not different between the two groups. Repeat ablation for the treatment of atrial arrhythmia recurrence was performed in 60 patients (SR: 44 [30%] patients; AF: 16 [33%] patients), with a trend towards a lower rate of pulmonary vein reconnections in the AF group (p = 0.07). There was no difference in 3-year arrhythmia-free survival (p = 0.8).

Conclusions: Cryoballoon-based PVI during AF results in lower nadir balloon temperatures and a trend towards a higher durability of PVI as compared to procedures performed in SR. The rate of real-time PVI recordings was not affected by the intraprocedural heart rhythm. (Cardiol J 2022; 29, 5: 807–814)

Key words: clinical outcomes, cryoballoon ablation, pulmonary vein isolation, intraprocedural heart rhythm

Introduction

Cryoballoon (CB) ablation has become an established treatment option for patients with atrial fibrillation (AF) [1–4]. The procedural endpoint is electrical isolation of the pulmonary veins (PV) [5]. Recent data indicate improved durability of pulmonary vein isolation (PVI) after CB ablation than after radiofrequency current (RFC) ablation [6]. Nevertheless, the number of patients requiring repeat ablation because of recurrences of AF — mainly driven by reconnected PV — remains high following either approach [7]. In contrast to point-by-point RFC ablation, the CB does not provide different energy levels, e.g., does not allow the use of higher energy settings at locations of poor tissue-catheter contact, such as the ridge between the left PV and the left atrial appendage. Once the CB has tissue contact at the antral aspect of the targeted PV, cryothermal energy is delivered via the entire distal hemisphere of the balloon. Durable, transmural lesion formation depends on the freeze cycle duration, and, more importantly, on the tissue contact of the CB. Since the latter cannot be measured directly, complete occlusion of the PV, duration of the temperature drop and nadir CB temperature serve as important surrogate parameters [8]. Theoretically, the reduced atrial contractility during AF might result in an enhanced stability of the balloon catheter and thereby, a more effective freeze cycle. However, to date, the impact of the intraprocedural rhythm (sinus rhythm [SR] vs. AF) during CB ablation has not yet been investigated.

The aim of the present study was to assess acute procedural characteristics, durability of PVI in patients with repeat ablation procedures and long-term clinical outcomes of CB ablation performed during AF or in SR.

Methods

Inclusion and exclusion criteria

Patients with symptomatic paroxysmal or persistent AF (as defined by 2016 ESC guidelines [5]) were included in the current study. Exclusion criteria were prior left atrial (LA) ablation procedures, a LA diameter > 60 mm, severe valvular heart disease or contraindications to post-interventional oral anticoagulation. Procedures were performed at two centers (Asklepios Klinik St. Georg, Hamburg, Germany, and German Heart Institute, Berlin, Germany).

The current study constitutes a retrospective analysis based oninstitutional databases. The study was approved by the local ethics boards and performed in accordance with the Declaration of Helsinki of 2013.

Preprocedural management

Transesophageal echocardiography was performed prior to PVI in all patients to rule out intracardiac thrombi and to assess LA diameter. No further pre-procedural imaging was performed.

Intraprocedural management

The intraprocedural management has been described in detail before [4, 9, 10]. In brief, in patients on vitamin K antagonists the procedure was performed under therapeutic international normalized ratio (INR) values of 2–3. Novel oral anticoagulants were stopped the day before the procedure and were later resumed 6 hours post ablation. All procedures were performed under deep sedation using midazolam, sufentanyl and propofol. One or two diagnostic catheters were introduced via the femoral vein and/or the left subclavian vein and were positioned within the coronary sinus and/or along the His bundle. A single transseptal puncture was performed via the femoral vein under fluoroscopic guidance, using a modified Brockenbrough technique and an 8.5 French (F) transseptal sheath (SL1, St. Jude Medical Inc., St. Paul, USA). After transseptal puncture, heparin boluses were administered in 30-minute intervals targeting an activated clotting time of ≥ 300 s. Selective angiographic visualization by dye injections or rotational angiography was performed to identify the individual PV ostia.

Cryoballoon-based PVI

The 28-mm second-generation CB was utilized exclusively in this study. The transseptal sheath was exchanged over a guidewire for a 12 F steerable sheath (FlexCath AdvanceTM, Medtronic, Inc.), through which the CB was advanced into the LA. Guiding of the CB to the target PV was performed over a 20-mm inner-lumen circular mapping catheter (Achieve™, Medtronic, Inc.) and complete occlusion of the PV ostium was verified by contrast injection through the central lumen of the inflated CB.

Patients were treated based on a “time-to-isolation” guided ablation protocol, i.e., after realtime verification of PVI, freezing was continued for an additional 120 s. In cases where the “time-to- -isolation” could not be recorded, the freeze-cycle duration was set at 180 s; no additional bonusfreeze cycle was applied after successful PVI [9]. In patients undergoing CB ablation during AF, electrical cardioversion was performed at the end of the procedure to restore SR.

An esophageal temperature probe (Sensitherm, St Jude Medical, Inc.; or Circa, Circa Scientific, Inc.) was inserted and positioned according to the individual CB position to provide esophageal temperature monitoring. The intraluminal esophageal temperature cut-off was set at 15°C [11].

During CB ablation along the septal PV, continuous phrenic nerve (PN) pacing was performed using a diagnostic catheter positioned in the superior vena cava (6 F, InquiryTM, St. Jude Medical, Inc.). PN capture was monitored by tactile feedback of diaphragmatic contraction by placing the operator’s hand on the patient’s abdomen. In addition, the continuous motor action potential was monitored and delivery of the refrigerant was stopped immediately if weakening or loss of diaphragmatic movement was noted or the amplitude of the continuous motor action potential decreased by 30% [12, 13].

Postprocedural care

Transthoracic echocardiography was performed in all patients to rule out a pericardial effusion. All patients were treated with proton-pump inhibitors for 4–6 weeks. Low molecular-weight heparin was administered in patients on vitamin K antagonists and an INR < 2.0 until a therapeutic INR of 2–3 was reached. Anticoagulation was continued for at least 3 months, and thereafter based on the individual CHA2DS2-VASc score. Previously ineffective antiarrhythmic drugs were continued for 3 months.

Follow-up

Following a blanking period of 3 months, patients completed outpatient clinical visits at 3, 6 and 12 months and in 6-month intervals thereafter according to our institutional standard; the clinical visits included electrocardiograms (ECGs) and 24 h-Holter ECGs. In patients with an implantable cardiac device with rhythm monitoring, regular device interrogations were additionally performed in order to detect arrhythmia recurrences. Moreover, regular telephone interviews were performed. In case of symptoms suggestive of recurrent arrhythmia, patients were advised to contact the outpatient clinic and subsequent clinical visits were immediately initiated.

Endpoints

The primary endpoints of this investigation were acute procedural characteristics, i.e., nadir balloon temperatures, procedure durations, fluoroscopy times, rate of real-time PVI recordings, and reconnection of initially isolated PV (assessed in patients with repeat ablation for recurrence of AF). Secondary endpoints were a documented recurrence of atrial arrhythmia with a duration of > 30 s outside the 3-month blanking period and complications, e.g., transient ischemic attack, stroke, pericardial tamponade, PN paralysis (PNP), and severe bleeding requiring blood transfusion.

Statistical analysis

All data were evaluated retrospectively. Continuous data are described as mean and standard deviation (SD) if normally distributed; otherwise the median and interquartile range [IQR, first quartile; third quartile] are reported. Categorical data are described with absolute and relative frequencies. Based on a logistic regression model (global test of no regression) baseline variables were simultaneously compared between the two groups.

Differences in procedural data were analyzed with a linear mixed model. Freedom from atrial arrhythmia recurrence was estimated with the Kaplan-Meier method. Differences in recurrencefree survival were analyzed with the log-rank test. All p-values were two-sided and a p-value < 0.05 was considered statistically significant. All calculations were performed with the statistical analysis software R (R Core Team, 2019).

Results

Patients

A total of 195 patients with symptomatic paroxysmal (n = 136) or persistent AF (n = 59) underwent PVI with the 28-mm second-generation CB. Ablation procedures were either performed in SR (n = 147) or during AF (n = 48). Persistent AF was more prevalent in the AF group than in the SR group (62% vs. 20%). All other baseline patient characteristics did not differ between the two groups (Table 1).

Table 1. Baseline patient characteristics.

Characteristics

SR group (n = 147)

AF group (n = 48)

Age [years]

63 [54;70]

66 [55;71]

Male gender

92 (63)

33 (69)

Paroxysmal AF

118 (80)

18 (38)

Persistent AF

29 (20)

30 (62)

CHA2DS2-VASc score:

0

28 (19)

8 (17)

1

37 (25)

8 (17)

2

37 (25)

14 (29)

3

32 (22)

11 (23)

4

8 (5)

5 (10)

5

5 (3)

2 (4)

BMI [kg/m2]

28 ± 5

27 ± 4

LA diameter [mm]

42 ± 5

45 ± 5

LVEF < 50%

9 (6)

3 (6)

Coronary artery disease

17 (12)

6 (12)

Diabetes mellitus

17 (12)

5 (10)

COPD

9 (6)

2 (4)

Smoker

19 (13)

5 (10)

Arterial hypertension

91 (62)

29 (60)

Procedural characteristics

In 193/195 patients, all PV were successfully isolated, while in 2 patients of the SR group both right inferior PV and one right superior PV were not isolated due to the occurrence of PNP. The nadir temperature during the freeze cycle applications was significantly lower in patients treated during AF than in patients treated in SR (–49 [IQR, –44; –54]°C vs. –47 [–42; –52]°C, p = 0.002). Median procedure and fluoroscopy times as well as the rate of time-to-isolation recordings did not differ between the two cohorts. Procedural parameters are given in Table 2.

Table 2. Procedural data.

Characteristics

SR group (n = 147)

AF group (n = 48)

P

Total procedure time [min]

130 [110; 152]

135 [120; 145]

0.497

Fluoroscopy time [min]

18 [14; 25]

20 [16; 24]

0.364

Nadir balloon temperature [°C]:

–47 [–42; –52]

–49 [–44; –54]

0.002

RSPV

–50 [–44; –53]

–52 [–48; –55]

RIPV

–46 [–41; –49]

–48 [–43; –54]

LSPV

–47 [–42; –52]

–50 [–43; –56]

LIPV

–44 [–41; –49]

–48 [–43; –52]

LCPV

–55 [–47; –59]

–47 [–45; –51]

Real-time PVI recording:

182 (31)

59 (30)

0.950

RSPV

48 (33)

18 (38)

RIPV

40 (28)

10 (21)

LSPV

48 (34)

16 (33)

LIPV

41 (29)

14 (29)

LCPV

5 (38)

1 (25)

PV with electrical reconnection*:

100 (57)

26 (41)

0.070

RSPV

23 (52)

8 (50)

RIPV

28 (64)

10 (62)

LSPV

23 (55)

3 (19)

LIPV

24 (57)

5 (31)

LCPV

2 (100)

Periprocedural complications

Procedural complications were two cardiac tamponades requiring pericardiocentesis, an arteriovenous fistula and two groin hematomas with conservative treatment each. PNP occurred in 3 patients resulting in absence of PVI in two right inferior PV and one right superior PV as stated previously. In 1 patient all PV were successfully isolated at the time of PNP. No atrio-esophageal fistula, no symptomatic PV stenosis, no stroke and no procedure-related deaths were observed.

Repeat ablation procedures

Repeat ablation for the treatment of atrial arrhythmia recurrence following the index CB procedure was performed in 60 patients (SR group: 44 [30%] patients; AF group: 16 [33%] patients). Procedures were exclusively performed with the use of RFC guided by 3-dimensional mapping. All patients showed electrical reconnection of at least one initially isolated PV. All PV were successfully re-isolated. There was a trend towards a lower rate of PV reconnections in the AF group without being statistically significant (p = 0.07). Characteristics of PV reconnections during repeat ablation procedures are given in Table 2.

Clinical follow-up

In the SR and the AF group, 1-, 2- and 3-year estimated arrhythmia-free survival was 78% and 75%, 64% and 69%, and 57% and 57%, respectively. The log-rank test did not indicate a significant difference in arrhythmia-free survival between the groups (p = 0.8). Recurrence-free survival is bshown in Figure 1.

170982.png
Figure 1. Freedom from atrial arrhythmia recurrence after cryoballoon ablation performed in sinus rhythm (SR; orange curve) or during atrial fibrillation (AF; blue curve). The log-rank test revealed no significant difference in arrhythmia-free survival between the groups (p = 0.8).

Discussion

Main findings

According to available research, the current study represents the first analysis investigating the impact of the intraprocedural heart rhythm during CB-based PVI on procedural characteristics and clinical outcomes. The main findings are: 1) significant lower nadir CB temperatures in patients undergoing PVI during AF than in patients being treated in SR; 2) a trend towards improved durability of PVI performed during AF; and 3) no influence of the basic rhythm on procedure and fluoroscopy times as well as the rate of time-to-isolation recordings during PVI. The 3-year arrhythmia-free survival did not differ between the two groups.

Key factors on efficacious CB ablation

Over the last decade, CB ablation aiming at PVI has become an established treatment for AF. During this time, several predictors of freedom from recurrence of AF and permanent PVI have been identified. As such, the time to PVI (confirmed by real-time recordings), the thaw time, and the nadir temperature of the CB impact PVI durability [8, 14–16]. Each of these parameters represents a surrogate of the balloon-tissue contact and energy transfer, which is still the most relevant aspect in creating durable, transmural cryolesions. The present study provides new insights into the impact of the intraprocedural heart rhythm on procedural parameters and clinical outcomes of CB ablation. One important result is the significantly lower nadir CB temperature in the AF group indicating that the reduced atrial contractility during AF is beneficial for balloon-tissue contact and en- ergy transfer, respectively. The second-generation CB features a refrigerant injection system that provides homogeneous cooling of the entire distal hemisphere of the CB. Temperature monitoring of the CB is realized via a thermocouple located inside the balloon. Reduced atrial contractility during AF seems to enhance adherence of the CB to the atrial myocardium. As a result, the surrounding blood flow, which results in a competing warming of the balloon, is reduced. This allows for lower nadir CB temperatures and, thereby, for more effective CB applications. Accordingly, patients in the AF group appear to be at a lower risk for PV reconnection. Interestingly, data on the impact of the intraprocedural heart rhythm during point-by- -point RFC ablation are inconsistent [17, 18]. It is conceivable that atrial contractility might play a greater role in “single-shot” CB ablation. During the first phase of the freeze cycle adherence of the balloon to the myocardium is essential for effective energy transfer to the tissue. The reduced atrial contractility during AF appears to support this process. Point-by-point RFC ablation is more complex and anatomic locations, e.g., the anterior or posterior aspect of the PV as well as individual anatomic variations, might pose a greater impact as compared to the relatively simple use of the CB.

Another relevant finding is that real-time recordings and proof of the occurrence of PVI is not hampered by performing CB ablation during AF. Real-time assessment of PVI is essential when applying time-to-isolation guided ablation protocols that pave the way for shorter procedure times without compromising procedural efficacy and safety [9, 19]. Aryana et al. [8] demonstrated that isolation of the PV within the first 60 s of the freeze-cycle is associated with a high durability of PVI. Therefore, the time to isolation provides one of the most powerful predictors of effective CB ablation [8].

Arrhythmia free-survival

In the current study, arrhythmia-free survival was not different between patients undergoing CB-based PVI during AF or in SR. A low nadir CB temperature is associated with a very high likelihood of durable PVI; however, our findings noted a significantly lower nadir CB temperature and a trend towards a higher durability of PVI in the AF group did not result in improved arrhythmia-free survival. The underlying causes might be multifactorial. First, a higher number of patients than provided in our analysis might be needed to demonstrate improved freedom from AF. Second, the prevalence of persistent AF was markedly higher in the AF group, and, as the experiences of previous studies have shown, results of catheter ablation for persistent AF are less successful than for paroxysmal AF [20–22]. Taking this into account, the similar arrhythmia-free survival in both groups is a positive result for the AF group.

Noteworthy, none of the 60 patients undergoing repeat ablation of AF showed permanent isolation of all PV underlining the impact of recovered PV conduction as one of the main drivers for recurrence of AF following PVI [23]. Yet, the rate of durable PVI was unexpected low at the time of repeat ablation, since previous studies reported on a proportion of patients with permanent isolation of all PV following CB ablation of 15% [6], 22% [24], and 53% [25]. However, even if these data were more convincing, durability of PVI is still not yet satisfactory. The implementation of the CB has led to greater availability of AF ablation, better reproducibility of clinical outcomes [26], and improved lesion characteristics as compared to RFC ablation [27], but the high proportion of reconnected PV demands further efforts ensuring permanent PVI following the index procedure. It also needs to be taken into consideration that only patients with symptomatic AF-recurrences were included in the study. It can be speculated that the proportion of durably isolated PVs is higher in patients without AF recurrences.

Limitations of the study

The current study is an observational, non-randomized analysis. The mode of the follow-up might have resulted in an overestimation of freedom from atrial arrhythmia recurrences. A sufficiently powered randomized comparison is mandatory before final conclusions can be drawn.

Conclusions

In this observational study, the performance of CB-based PVI during AF resulted in lower nadir CB temperatures and a trend towards higher durability of PVI when compared to procedures performed in SR. The rate of real-time PVI recording was not affected by the intraprocedural rhythm. In accordance to the present findings, in CB ablation, electrical cardioversion of AF should be performed once PVI has been obtained.

Conflict of interest: Andreas Metzner received speaker honoraria and travel grants from Medtronic. Andreas Rillig received travel grants from Biosense Webster, Hansen Medical, Medtronic, EPSolutions and St. Jude Medical and lecture fees from St. Jude Medical, Medtronic and Boehringer Ingelheim, consultant fees from Medtronic and took part at the Boston Scientific EP fellowship. Bruno Reissmann and Christian-H. Heeger received travel grants from Medtronic. Karl-Heinz Kuck received research grants from Medtronic and speaker bureau’s honoraria from Biosense Webster, Impulse Dynamics, and Biotronik. Doreen Schöppenthau received travel grants from St. Jude Medical, Bristol-Myers-Squibb and Biosense Webster, a research grant from Biosense Webster and took part in the Boston scientific EP fellowship program.

References

  1. Kuck KH, Brugada J, Fürnkranz A, et al. FIRE AND ICE Investigators. Cryoballoon or Radiofrequency Ablation for Paroxysmal Atrial Fibrillation. N Engl J Med. 2016; 374(23): 2235–2245, doi: 10.1056/NEJMoa1602014, indexed in Pubmed: 27042964.
  2. Kuck KH, Fürnkranz A, Chun K, et al. Cryoballoon or radiofrequency ablation for symptomatic paroxysmal atrial fibrillation: reintervention, rehospitalization, and quality-of-life outcomes in the FIRE AND ICE trial. Eur Heart J. 2016; 37(38): 2858–2865, doi: 10.1093/eurheartj/ehw285.
  3. Luik A, Radzewitz A, Kieser M, et al. Cryoballoon versus open irrigated radiofrequency ablation in patients with paroxysmal atrial fibrillation: the prospective, randomized, controlled, noinferiority freezeaf study. Circulation. 2015; 132(14): 1311–1319, doi: 10.1161/CIRCULATIONAHA.115.016871, indexed in Pubmed: 26283655.
  4. Metzner A, Reissmann B, Rausch P, et al. One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circ Arrhythm Electrophysiol. 2014; 7(2): 288–292, doi: 10.1161/CIRCEP.114.001473, indexed in Pubmed: 24610797.
  5. Kirchhof P, Benussi S, Kotecha D, et al. 2016 ESC Guidelines for the management of atrial fibrillation developed in collaborationn with EACTS: The Task Force for the management of atrial fibrillation of the European Society of Cardiology (ESC). Developed with the special contribution of the European Heart Rhythm Association (EHRA) of the ESCEndorsed by the European Stroke Organisation (ESO). Eur Heart J. 2016; 37: 2893–2962.
  6. Kuck KH, Albenque JP, Chun K, et al. Repeat ablation for atrial fibrillation recurrence post cryoballoon or radiofrequency ablation in the FIRE AND ICE trial. Circ Arrhythm Electrophysiol. 2019; 12(6), doi: 10.1161/circep.119.007247.
  7. Inaba O, Metzner A, Rottner L, et al. Radiofrequency or cryoballoon ablation for index pulmonary vein isolation: What is the impact on long-term clinical outcomes after repeat ablation? J Cardiovasc Electrophysiol. 2020; 31(5): 1068–1074, doi: 10.1111/jce.14432, indexed in Pubmed: 32128924.
  8. Aryana A, Mugnai G, Singh SM, et al. Procedural and biophysical indicators of durable pulmonary vein isolation during cryoballoon ablation of atrial fibrillation. Heart Rhythm. 2016; 13(2): 424–432, doi: 10.1016/j.hrthm.2015.10.033, indexed in Pubmed: 26520204.
  9. Reissmann B, Wissner E, Deiss S, et al. First insights into cryoballoon-based pulmonary vein isolation taking the individual timeto-isolation into account. Europace. 2017; 19(10): 1676–1680, doi: 10.1093/europace/euw233, indexed in Pubmed: 28201538.
  10. Heeger CH, Wissner E, Wohlmuth P, et al. Bonus-freeze: benefit or risk? Two-year outcome and procedural comparison of a „bonus-freeze” and „no bonus-freeze” protocol using the second- generation cryoballoon for pulmonary vein isolation. Clin Res Cardiol. 2016; 105(9): 774–782, doi: 10.1007/s00392-016-0987-8, indexed in Pubmed: 27085722.
  11. Fürnkranz A, Bordignon S, Böhmig M, et al. Reduced incidence of esophageal lesions by luminal esophageal temperature-guided second-generation cryoballoon ablation. Heart Rhythm. 2015; 12(2): 268–274, doi: 10.1016/j.hrthm.2014.10.033, indexed in Pubmed: 25446159.
  12. Metzner A, Rausch P, Lemes C, et al. The incidence of phrenic nerve injury during pulmonary vein isolation using the se ondgeneration 28 mm cryoballoon. J Cardiovasc Electrophysiol. 2014; 25(5): 466–470, doi: 10.1111/jce.12358, indexed in Pubmed: 24400647.
  13. Mondésert B, Andrade JG, Khairy P, et al. Clinical experience with a novel electromyographic approach to preventing phrenic nerve injury during cryoballoon ablation in atrial fibrillation. Circ Arrhythm Electrophysiol. 2014; 7(4): 605–611, doi: 10.1161/ CIRCEP.113.001238, indexed in Pubmed: 25017398.
  14. Chierchia GB, de Asmundis C, Namdar M, et al. Pulmonary vein isolation during cryoballoon ablation using the novel Achieve inner lumen mapping catheter: a feasibility study. Europace. 2012; 14(7): 962–967, doi: 10.1093/europace/eus041, indexed in Pubmed: 22411731.
  15. Ghosh J, Martin A, Keech AC, et al. Balloon warming time is the strongest predictor of late pulmonary vein electricalmreconnection following cryoballoon ablation for atrial fibrillation. Heart Rhythm. 2013; 10(9): 1311–1317, doi: 10.1016/j. hrthm.2013.06.014, indexed in Pubmed: 23792110.
  16. Reissmann B, Plenge T, Heeger CH, et al. Predictors of freedom from atrial arrhythmia recurrence after cryoballoon ablation for persistent atrial fibrillation: a multicenter study. J Cardiovasc Electrophysiol. 2019; 30(9): 1436–1442, doi: 10.1111/jce.14023, indexed in Pubmed: 31190440.
  17. Matsuda H, Parwani AS, Attanasio P, et al. Atrial rhythm influences catheter tissue contact during radiofrequency catheter ablation of atrial fibrillation: comparison of contact force between sinus rhythm and atrial fibrillation. Heart Vessels. 2016; 31(9): 1544–1552, doi: 10.1007/s00380-015-0763-0, indexed in Pubmed: 26498938.
  18. Sarkozy A, Shah D, Saenen J, et al. Contact force in atrial fibrillation: role of atrial rhythm and ventricular contractions: co-force atrial fibrillation study. Circ Arrhythm Electrophysiol. 2015; 8(6): 1342–1350, doi: 10.1161/CIRCEP.115.003041, indexed in Pubmed: 26383774.
  19. Chun KR, Stich M, Fürnkranz A, et al. Individualized cryoballoon energy pulmonary vein isolation guided by real-time pulmonary vein recordings, the randomized ICE-T trial. Heart Rhythm. 2017; 14(4): 495–500, doi: 10.1016/j.hrthm.2016.12.014, indexed in Pubmed: 27956248.
  20. Reissmann B, Plenge T, Heeger CH, et al. Predictors of freedom from atrial arrhythmia recurrence after cryoballoon ablation for persistent atrial fibrillation: a multicenter study. J Cardiovasc Electrophysiol. 2019; 30(9): 1436–1442, doi: 10.1111/jce.14023, indexed in Pubmed: 31190440.
  21. Scherr D, Khairy P, Miyazaki S, et al. Five-year outcome of catheter ablation of persistent atrial fibrillation using termination of atrial fibrillation as a procedural endpoint. Circ Arrhythm Electrophysiol. 2015; 8(1): 18–24, doi: 10.1161/CIRCEP.114.001943, indexed in Pubmed: 25528745.
  22. Schreiber D, Rostock T, Fröhlich M, et al. Five-year follow-up after catheter ablation of persistent atrial fibrillation using the stepwise approach and prognostic factors for success. Circ Arhythm Electrophysiol. 2015; 8(2): 308–317, doi: 10.1161/CIRCEP.114.001672, indexed in Pubmed: 25744570.
  23. Ouyang F, Antz M, Ernst S, et al. Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation. 2005; 111(2): 127–135, doi: 10.1161/01.CIR.0000151289.73085.36, indexed in Pubmed: 15623542.
  24. Buist TJ, Adiyaman A, Smit JJ, et al. Arrhythmia-free survival and pulmonary vein reconnection patterns after second-generation cryoballoon and contact-force radiofrequency pulmonary vein isolation. Clin Res Cardiol. 2018; 107(6): 498–506, doi: 10.1007/s00392-018-1211-9, indexed in Pubmed: 29411114.
  25. Aryana A, Singh SM, Mugnai G, et al. Pulmonary vein reconnection following catheter ablation of atrial fibrillation using the second-generation cryoballoon versus open-irrigated radiofrequency: results of a multicenter analysis. J Interv Card Electrophysiol. 2016; 47(3): 341–348, doi: 10.1007/s10840-016-0172-z, indexed in Pubmed: 27475949.
  26. Providencia R, Defaye P, Lambiase PD, et al. Results from a multicentre comparison of cryoballoon vs. radiofrequency ablation for paroxysmal atrial fibrillation: is cryoablation more reproducible? Europace. 2017; 19(1): 48–57, doi: 10.1093/europace/ euw080, indexed in Pubmed: 27267554.
  27. Kurose J, Kiuchi K, Fukuzawa K, et al. The lesion characteristics assessed by LGE-MRI after the cryoballoon ablation and conventional radiofrequency ablation. J Arrhythm. 2018; 34(2): 158–166, doi: 10.1002/joa3.12025, indexed in Pubmed: 29657591.