Implantation of the Micra transcatheter pacing system: Single Polish center experience with the real costs of hospitalization analysis

Marcin Grabowski1, Marcin Michalak1, Monika Gawałko1, Sylwia Gajda1, Andrzej Cacko1, Łukasz Januszkiewicz1, Agnieszka Kołodzińska1, Przmysław P. Mitkowski2, Gabor Z. Duray3, Grzegorz Opolski1
Pubmed: 30155871
Cardiol J 2020;27(1):47-53.

Abstract

Background: The Micra transcatheter pacing system (TPS) is a miniaturized, single-chamber pacemaker system. Study reported herein is an initial experience with implantation of the Micra TPS.

Methods: The leadless pacemaker was implanted in 10 patients with standard indications for a permanent pacemaker implantation. All hospitalization costs were calculated for all patients.

Results: The mean age of the patients was 75 ± 7.1 years, 6 were men and 4 were women. Four patients had permanent atrial fibrillation as the basal rhythm and 6 patients had sinus rhythm. All patients had at least one relative contraindication that precluded the use of a traditional pacing system. Mean intraoperative ventricular sensing amplitude was 10.6 ± 5.4 mV, impedance 843 ± 185 ohms, and pacing threshold at 0.24 ms was 0.56 ± 0.23 V. At discharge, those values were 13.9 ± 5.6 mV, 667 ± 119 ohms and 0.47 ± 0.17, respectively. The mean duration of implantation procedure was 82 min, while mean fluoroscopy time was 3.5 min. Two patients developed hematoma at the groin puncture site post-implantation. In 1 case there was a need for erythrocyte mass transfusion and surgical intervention. Mean total time of hospitalization was 26 days and time from procedure to discharge 12 days. Average cost of hospitalization per 1 patient was 11,260.15 EUR minimal cost was 9,052.68 EUR, while maximal cost was 16,533.18 EUR.

Conclusions: Implantation of leadless pacemakers is feasible, safe and provides advantages over the conventional system. Hospitalization costs vary for individual patients in wide range.

Article available in PDF format

View PDF Download PDF file

References

  1. Udo EO, Zuithoff NPA, van Hemel NM, et al. Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm. 2012; 9(5): 728–735.
  2. Kirkfeldt RE, Johansen JB, Nohr EA, et al. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014; 35(18): 1186–1194.
  3. Ritter P, Duray GZ, Steinwender C, et al. Early performance of a miniaturized leadless cardiac pacemaker: the Micra Transcatheter Pacing Study. Eur Heart J. 2015; 36(37): 2510–2519.
  4. Knops RE, Tjong FVY, Neuzil P, et al. Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial. J Am Coll Cardiol. 2015; 65(15): 1497–1504.
  5. El-Chami MF, Roberts PR, Kypta A, et al. How to Implant a Leadless Pacemaker With a Tine-Based Fixation. J Cardiovasc Electrophysiol. 2016; 27(12): 1495–1501.
  6. Sohail MR, Hussain S, Le KY, et al. Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections. J Am Coll Cardiol. 2007; 49(18): 1851–1859.
  7. Baddour LM, Epstein AE, Erickson CC, et al. Update on cardiovascular implantable electronic device infections and their management: a scientific statement from the American Heart Association. Circulation. 2010; 121(3): 458–477.
  8. Kypta A, Blessberger H, Kammler J, et al. Leadless cardiac pacemaker implantation after lead extraction in patients with severe device infection. J Cardiovasc Electrophysiol. 2016; 27(9): 1067–1071.
  9. Reynolds D, Duray G, Omar R, et al. A Leadless Intracardiac Transcatheter Pacing System. N Engl J Med. 2016; 374(6): 533–541.
  10. Roberts PR, Clementy N, Al Samadi F, et al. A leadless pacemaker in the real-world setting: The Micra Transcatheter Pacing System Post-Approval Registry. Heart Rhythm. 2017; 14(9): 1375–1379.
  11. Reddy VY, Exner DV, Cantillon DJ, et al. Percutaneous Implantation of an Entirely Intracardiac Leadless Pacemaker. N Engl J Med. 2015; 373(12): 1125–1135.
  12. Duray GZ, Ritter P, El-Chami M, et al. Long-term performance of a transcatheter pacing system: 12-Month results from the Micra Transcatheter Pacing Study. Heart Rhythm. 2017; 14(5): 702–709.