Conduction system engagement by mid-septal leadless pacemaker in a patient with persistent iatrogenic atrioventricular block

Giusy Sirico1Massimo Mantica1Domenico Sirico2Andrea Montisci34Federica Calabrese5Massimo Pala5Daniele Malaspina5
1Department of Cardiac Electrophysiology and Pacing, Istituto Clinico Sant’Ambrogio, Milan, Italy
2Pediatric and Congenital Cardiology Unit, Department of Woman and Child’s Health, University of Padova, Italy
3Department of Anesthesia and Intensive Care, Istituto Clinico Sant’Ambrogio, Milan, Italy
4Chair of Cardiac Surgery, Postgraduate in Cardiac Surgery, University of Milan, Italy
5Department of Cardiology and Pacing, ASST Santi Paolo e Carlo, Milan, Italy

Address for correspondence: Giusy Sirico, MD, PhD, Via L.G.Faravelli n.16, Istituto Clinico Sant’Ambrogio, 20149, Milan, Italy, e-mail: giusy.sirico@gmail.com

Received: 7.04.2021 Accepted: 22.08.2021

This article is available in open access under Creative Common Attribution-Non-Commercial-No Derivatives 4.0 International (CC BY-NC-ND 4.0) license, allowing to download articles and share them with others as long as they credit the authors and the publisher, but without permission to change them in any way or use them commercially.

An 89-year-old woman was admitted to our Arrhythmology Unit for syncope. Electrocardiogram (ECG) showed sinus rhythm with left bundle branch block (LBBB) (Fig. 1A). Holter ECG monitoring showed paroxysmal atrial fibrillation with difficult rate control. The patient was referred for ‘ablate and pace’ strategy. Left subclavian vein occlusion prevented the traditional pacing implant. Micra™ TPS was implanted in the right ventricle mid-septal position causing right bundle branch (RBB) bump and atrioventricular block that was promptly treated with a temporary pacing wire (Fig. 1B). Micra was successfully deployed (Fig. 1C, D). Pacing threshold was 0.50 V @ 0.24 ms, sensing was 6.4 mV and impedance was 610 Ohm. Pacemaker dependency was persistent. Paced QRS was identical in morphology and duration to baseline QRS: duration of 130 ms, rS complex in inferior leads, positive R wave in lead I and aVL with a typical notch at the onset of QRS (Fig. 1E). Pa cing at high output (5 Volt @ 1 ms) did not change QRS morphology (Fig. 1F). There are two possible explanations of this phenomenon: 1) the different localizations between the site of RBB induced block and Micra™ pacing site, with potential selective activation of the right bundle; 2) the greater surface of endocardial contact of the Micra resulted in the capture of a greater portion of conduction tissue fibers connected downstream from the site of RBB block. The ability of Micra leadless pacemakers to selectively pace the conduction system could allow remarkable advantages in terms of QRS duration.

Sirico_01_poprawiony.tif

Figure 1. Micra™ implantation and paced electrocardiogram (ECG); A. Baseline 12 lead ECG at admission; B. Right anterior oblique view of Micra™ implantation and temporary pacing lead; C, D. Right and left anterior oblique view of Micra™ system successfully deployed at right ventricle mid-septum; E. 12 lead ECG after Micra™ implantation with pacing at 2.5 V @ 0.4 ms; F. 12 lead ECG with pacing at 5 V @ 1 ms. The panels B, C and D show the mechanical valve 27 SJM Regent previously implanted in the mitral position.

Conflict of interest: None declared