Continuous hands-free monitoring of echocardiographic exercise test using probe fixation device

Katarzyna Wdowiak-Okrojek1, Piotr Lipiec1, Paulina Wejner-Mik1, Zbigniew Bednarkiewicz1, Jarosław D. Kasprzak1
Affiliations
  1. Student Society at 1st Department of Cardiology, Medical University of Lodz, Łódź, Poland

open access

Ahead of print
Original articles
Submitted: 2021-06-10
Accepted: 2022-03-24
Published online: 2022-05-13

Abstract

Background: Stress echocardiography has been widely used in clinical practice for decades and has recently gained even more importance in diagnostic approaches to ischemic heart disease. However, it still has numerous limitations. Despite advantages of physical exercise as most physiologic stressor, it is impossible to continuously monitor the cardiac function during treadmill test and difficult to maintain an optimal acoustic window during cycle ergometer exercise tests. The aim herein, is to assess the feasibility of probe fixation for use during exercise echocardiography. Methods: Forty-eight subjects (47 men, mean age 42 ± 17 years, 25 healthy volunteers, 23 patients with suspected coronary artery disease) were included in this study. All subjects underwent exercise stress test on treadmill (32 cases) or cycle ergometer (16 cases). Both sector and matrix probes were used (in 17 and 31 tests, respectively). The semi-quantitative quality of acquired apical views were assessed at each stage using a four-point grading system. Results: The mean time required for probe fixation was 9 ± 2 min. At baseline, 10 patients had at least one apical window of quality precluding reliable analysis. Twenty-five patients required probe repositioning during exercise (more often on a treadmill). During peak exercise quality of images in all views declined, but for diagnostic purposes it remained sufficient in 29 patients. Thus, 76% of performed tests (60% study population) had sufficient image quality. Conclusions: Probe fixation offers the possibility of continuous acquisition of echocardiographic images during physical exercise. The device is suitable almost exclusively for male patients and in some patients requires repositioning.

Abstract

Background: Stress echocardiography has been widely used in clinical practice for decades and has recently gained even more importance in diagnostic approaches to ischemic heart disease. However, it still has numerous limitations. Despite advantages of physical exercise as most physiologic stressor, it is impossible to continuously monitor the cardiac function during treadmill test and difficult to maintain an optimal acoustic window during cycle ergometer exercise tests. The aim herein, is to assess the feasibility of probe fixation for use during exercise echocardiography. Methods: Forty-eight subjects (47 men, mean age 42 ± 17 years, 25 healthy volunteers, 23 patients with suspected coronary artery disease) were included in this study. All subjects underwent exercise stress test on treadmill (32 cases) or cycle ergometer (16 cases). Both sector and matrix probes were used (in 17 and 31 tests, respectively). The semi-quantitative quality of acquired apical views were assessed at each stage using a four-point grading system. Results: The mean time required for probe fixation was 9 ± 2 min. At baseline, 10 patients had at least one apical window of quality precluding reliable analysis. Twenty-five patients required probe repositioning during exercise (more often on a treadmill). During peak exercise quality of images in all views declined, but for diagnostic purposes it remained sufficient in 29 patients. Thus, 76% of performed tests (60% study population) had sufficient image quality. Conclusions: Probe fixation offers the possibility of continuous acquisition of echocardiographic images during physical exercise. The device is suitable almost exclusively for male patients and in some patients requires repositioning.

Get Citation

Keywords

stress echocardiography, probe fixation, continuous monitoring

About this article
Title

Continuous hands-free monitoring of echocardiographic exercise test using probe fixation device

Journal

Cardiology Journal

Issue

Ahead of print

Article type

Original Article

Published online

2022-05-13

Page views

957

Article views/downloads

215

DOI

10.5603/CJ.a2022.0032

Pubmed

35578760

Keywords

stress echocardiography
probe fixation
continuous monitoring

Authors

Katarzyna Wdowiak-Okrojek
Piotr Lipiec
Paulina Wejner-Mik
Zbigniew Bednarkiewicz
Jarosław D. Kasprzak

References (22)
  1. Sicari R, Nihoyannopoulos P, Evangelista A, et al. Stress echocardiography expert consensus statement — executive summary: European Association of Echocardiography (EAE) (a registered branch of the ESC). Eur Heart J. 2009; 30(3): 278–289.
  2. Neumann FJ, Sousa-Uva M, Ahlsson A, et al. 2018 ESC/EACTS Guidelines on myocardial revascularization. Eur Heart J. 2019; 40(2): 87–165.
  3. Knuuti J, Wijns W, Saraste A, et al. 2019 ESC Guidelines for the diagnosis and management of chronic coronary syndromes. Eur Heart J. 2020; 41(3): 407–477.
  4. Lancellotti P, Dulgheru R, Go YY, et al. Stress echocardiography in patients with native valvular heart disease. Heart. 2018; 104(10): 807–813.
  5. Belyavskiy E, Morris DA, Url-Michitsch M, et al. Diastolic stress test echocardiography in patients with suspected heart failure with preserved ejection fraction: a pilot study. ESC Heart Fail. 2019; 6(1): 146–153.
  6. Pieske B, Tschöpe C, de Boer RA, et al. How to diagnose heart failure with preserved ejection fraction: the HFA-PEFF diagnostic algorithm: a consensus recommendation from the Heart Failure Association (HFA) of the European Society of Cardiology (ESC). Eur Heart J. 2019; 40(40): 3297–3317.
  7. Peteiro J, Bouzas-Mosquera A, Estevez R, et al. Head-to-head comparison of peak supine bicycle exercise echocardiography and treadmill exercise echocardiography at peak and at post-exercise for the detection of coronary artery disease. J Am Soc Echocardiogr. 2012; 25(3): 319–326.
  8. Badruddin SM, Ahmad A, Mickelson J, et al. Supine bicycle versus post-treadmill exercise echocardiography in the detection of myocardial ischemia: a randomized single-blind crossover trial. J Am Coll Cardiol. 1999; 33(6): 1485–1490.
  9. Blans MJ, Bosch FH, van der Hoeven JG. The use of an external ultrasound fixator (Probefix) on intensive care patients: a feasibility study. Ultrasound J. 2019; 11(1): 26.
  10. Banerjee A, Newman DR, Van den Bruel A, et al. Diagnostic accuracy of exercise stress testing for coronary artery disease: a systematic review and meta-analysis of prospective studies. Int J Clin Pract. 2012; 66(5): 477–492.
  11. Ryan T, Segar DS, Sawada SG, et al. Detection of coronary artery disease with upright bicycle exercise echocardiography. J Am Soc Echocardiogr. 1993; 6(2): 186–197.
  12. Płońska−Gościniak E, Gackowski A, Gąsior Z, et al. Recommendations of the Echocardiography Working Group of the Polish Cardiac Society for stress echocardiography use in clinical practice 2011. Kardiol Pol. 2011; 69(6): 642–648.
  13. Sicari R. Stress echocardiography: time for a reassessment? Int J Cardiol. 2018; 259: 47–48.
  14. Suzuki K, Hirano Y, Yamada H, et al. Practical guidance for the implementation of stress echocardiography. J Echocardiogr. 2018; 16(3): 105–129.
  15. Hecht HS, DeBord L, Sotomayor N, et al. Supine bicycle stress echocardiography: peak exercise imaging is superior to postexercise imaging. J Am Soc Echocardiogr. 1993; 6(3 Pt 1): 265–271.
  16. Caiati C, Lepera ME, Carretta D, et al. Head-to-head comparison of peak upright bicycle and post-treadmill echocardiography in detecting coronary artery disease: a randomized, single-blind crossover study. J Am Soc Echocardiogr. 2013; 26(12): 1434–1443.
  17. Chandraratna PAN, Gajanayaka R, Makkena SM, et al. "Hands-Free" continuous echocardiography during treadmill exercise using a novel ultrasound transducer. Echocardiography. 2010; 27(5): 563–566.
  18. Chandraratna PA, Vijayasekaran S, Brar R, et al. Feasibility of continuous transthoracic cardiac imaging using a novel ultrasound transducer. Echocardiography. 2001; 18(8): 651–655.
  19. Nakashiki K, Kisanuki A, Otsuji Y, et al. Usefulness of a novel ultrasound transducer for continuous monitoring treadmill exercise echocardiography to assess coronary artery disease. Circ J. 2006; 70(10): 1297–1302.
  20. Salden OAE, van Everdingen WM, Spee R, et al. How I do it: feasibility of a new ultrasound probe fixator to facilitate high quality stress echocardiography. Cardiovasc Ultrasound. 2018; 16(1): 6.
  21. Blans MJ, Bosch FH, van der Hoeven JG. The use of an external ultrasound fixator (Probefix) on intensive care patients: a feasibility study. Ultrasound J. 2019; 11(1): 26.
  22. Bouwmeester S, de Kleijn M, van Wijngaarden J, et al. The use of a probe stabilizer to reduce musculoskeletal overload of ultrasound operators in routine diagnostic echocardiographic imaging. J Ultrason. 2019; 19(78): 193–197.

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Grupa Via Medica, ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: viamedica@viamedica.pl