Transfermoral aortic valve implantation using self-expanding New Valve Technology (NVT) Allegra bioprosthesis: A pilot prospective study

Dariusz Jagielak, Aleksandra Stanska, Andrzej Klapkowski, Maciej Brzezinski, Maciej Kowalik, Dariusz Ciecwierz, Milosz Jaguszewski, Marcin Fijalkowski
DOI: 10.5603/CJ.a2019.0019
·
Pubmed: 30761515

open access

Ahead of print
Original articles
Published online: 2019-02-12

Abstract

Background: Transcatheter aortic valve implantation (TAVI) has become a standard therapeutic option for patients with severe aortic stenosis (AS) at high cardiac surgical risk. The aim of the NAUTILUS study was to investigate the safety and performance of the New Valve Technology (NVT) Allegra bioprosthesis in high-risk patients undergoing TAVI.

Methods: 27 patients with severe, symptomatic AS at high surgical risk were prospectively enrolled, who underwent treatment using the novel self-expanding NVT Allegra bioprosthesis via transfemoral approach (TF-TAVI). The primary end-point was all-cause mortality at 30 days.

Results: Patients were elderly (83 years, range 75–89 years), and predominantly female (70.4%, n = 19). All patients were deemed to be at high surgical risk, with a mean logistic EuroSCORE of 12.4% (range, 2.8–31.8%). The bioprosthesis was successfully implanted in 96% of the cases (n = 25). The echocardiographic assessment confirmed good hemodynamic profile after implantation of the NVT Allegra bioprosthesis. Complications included cardiac tamponade (4%, n = 1) and the need for permanent pacemaker implantation (8%, n = 2). The analysis of procedural aspects showed a short learning effect related to the precise placement of the valve. A significant improvement in clinical symptoms were observed, and no patients died in-hospital or within 30 days of post-discharge observation.

Conclusions: This prospective observation shows that the NVT Allegra bioprosthesis was associated with a satisfactory safety profile and a remarkable hemodynamic performance after implantation.

Abstract

Background: Transcatheter aortic valve implantation (TAVI) has become a standard therapeutic option for patients with severe aortic stenosis (AS) at high cardiac surgical risk. The aim of the NAUTILUS study was to investigate the safety and performance of the New Valve Technology (NVT) Allegra bioprosthesis in high-risk patients undergoing TAVI.

Methods: 27 patients with severe, symptomatic AS at high surgical risk were prospectively enrolled, who underwent treatment using the novel self-expanding NVT Allegra bioprosthesis via transfemoral approach (TF-TAVI). The primary end-point was all-cause mortality at 30 days.

Results: Patients were elderly (83 years, range 75–89 years), and predominantly female (70.4%, n = 19). All patients were deemed to be at high surgical risk, with a mean logistic EuroSCORE of 12.4% (range, 2.8–31.8%). The bioprosthesis was successfully implanted in 96% of the cases (n = 25). The echocardiographic assessment confirmed good hemodynamic profile after implantation of the NVT Allegra bioprosthesis. Complications included cardiac tamponade (4%, n = 1) and the need for permanent pacemaker implantation (8%, n = 2). The analysis of procedural aspects showed a short learning effect related to the precise placement of the valve. A significant improvement in clinical symptoms were observed, and no patients died in-hospital or within 30 days of post-discharge observation.

Conclusions: This prospective observation shows that the NVT Allegra bioprosthesis was associated with a satisfactory safety profile and a remarkable hemodynamic performance after implantation.

Get Citation

Keywords

aortic stenosis, clinical trial, elderly population, transcatheter aortic valve implantation

About this article
Title

Transfermoral aortic valve implantation using self-expanding New Valve Technology (NVT) Allegra bioprosthesis: A pilot prospective study

Journal

Cardiology Journal

Issue

Ahead of print

Article type

Research paper

Published online

2019-02-12

DOI

10.5603/CJ.a2019.0019

Pubmed

30761515

Keywords

aortic stenosis
clinical trial
elderly population
transcatheter aortic valve implantation

Authors

Dariusz Jagielak
Aleksandra Stanska
Andrzej Klapkowski
Maciej Brzezinski
Maciej Kowalik
Dariusz Ciecwierz
Milosz Jaguszewski
Marcin Fijalkowski

References (20)
  1. Aldalati O, MacCarthy P, Dworakowski R. Trans-catheter aortic valve implantation: Contemporary practice and the future. Cardiol J. 2017; 24(2): 206–215.
  2. Cribier A, Eltchaninoff H, Bash A, et al. Percutaneous transcatheter implantation of an aortic valve prosthesis for calcific aortic stenosis: first human case description. Circulation. 2002; 106(24): 3006–3008.
  3. Giordana F, D'Ascenzo F, Nijhoff F, et al. Meta-analysis of predictors of all-cause mortality after transcatheter aortic valve implantation. Am J Cardiol. 2014; 114(9): 1447–1455.
  4. Eggebrecht H, Mehta RH. Transcatheter aortic valve implantation (TAVI) in Germany 2008-2014: on its way to standard therapy for aortic valve stenosis in the elderly? EuroIntervention. 2016; 11(9): 1029–1033.
  5. Urena M, Webb JG, Eltchaninoff H, et al. Late cardiac death in patients undergoing transcatheter aortic valve replacement: incidence and predictors of advanced heart failure and sudden cardiac death. J Am Coll Cardiol. 2015; 65(5): 437–448.
  6. Ribeiro HB, Nombela-Franco L, Muñoz-García AJ, et al. Predictors and impact of myocardial injury after transcatheter aortic valve replacement: a multicenter registry. J Am Coll Cardiol. 2015; 66(19): 2075–2088.
  7. Gensas CS, Caixeta A, Siqueira D, et al. Predictors of permanent pacemaker requirement after transcatheter aortic valve implantation: insights from a Brazilian registry. Int J Cardiol. 2014; 175(2): 248–252.
  8. Wenaweser P, Stortecky S, Heg D, et al. Short-term clinical outcomes among patients undergoing transcatheter aortic valve implantation in Switzerland: the Swiss TAVI registry. EuroIntervention. 2014; 10(8): 982–989.
  9. Ribeiro HB, Webb JG, Makkar RR, et al. Predictive factors, management, and clinical outcomes of coronary obstruction following transcatheter aortic valve implantation: insights from a large multicenter registry. J Am Coll Cardiol. 2013; 62(17): 1552–1562.
  10. Athappan G, Patvardhan E, Tuzcu EM, et al. Incidence, predictors, and outcomes of aortic regurgitation after transcatheter aortic valve replacement: meta-analysis and systematic review of literature. J Am Coll Cardiol. 2013; 61(15): 1585–1595.
  11. Tchetche D, Van Mieghem NM. New-generation TAVI devices: description and specifications. EuroIntervention. 2014; 10 Suppl U: U90–U9U100.
  12. Wenaweser P, Stortecky S, Schütz T, et al. Transcatheter aortic valve implantation with the NVT Allegra transcatheter heart valve system: first-in-human experience with a novel self-expanding transcatheter heart valve. EuroIntervention. 2016; 12(1): 71–77.
  13. Leon MB, Piazza N, Nikolsky E, et al. Standardized endpoint definitions for Transcatheter Aortic Valve Implantation clinical trials: a consensus report from the Valve Academic Research Consortium. J Am Coll Cardiol. 2011; 57(3): 253–269.
  14. Siontis GCM, Jüni P, Pilgrim T, et al. Predictors of permanent pacemaker implantation in patients with severe aortic stenosis undergoing TAVR: a meta-analysis. J Am Coll Cardiol. 2014; 64(2): 129–140.
  15. Hayashida K, Morice MC, Chevalier B, et al. Sex-related differences in clinical presentation and outcome of transcatheter aortic valve implantation for severe aortic stenosis. J Am Coll Cardiol. 2012; 59(6): 566–571.
  16. Khawaja MZ, Rajani R, Cook A, et al. Permanent pacemaker insertion after CoreValve transcatheter aortic valve implantation: incidence and contributing factors (the UK CoreValve Collaborative). Circulation. 2011; 123(9): 951–960.
  17. Eggebrecht H, Schmermund A, Voigtländer T, et al. Risk of stroke after transcatheter aortic valve implantation (TAVI): a meta-analysis of 10,037 published patients. EuroIntervention. 2012; 8(1): 129–138.
  18. Chieffo A, Buchanan GL, Van Mieghem NM, et al. Transcatheter aortic valve implantation with the Edwards SAPIEN versus the Medtronic CoreValve Revalving system devices: a multicenter collaborative study: the PRAGMATIC Plus Initiative (Pooled-RotterdAm-Milano-Toulouse In Collaboration). J Am Coll Cardiol. 2013; 61(8): 830–836.
  19. Tchetche D, Farah B, Misuraca L, et al. Cerebrovascular events post-transcatheter aortic valve replacement in a large cohort of patients: a FRANCE-2 registry substudy. JACC Cardiovasc Interv. 2014; 7(10): 1138–1145.
  20. Stortecky S, Windecker S, Pilgrim T, et al. Cerebrovascular accidents complicating transcatheter aortic valve implantation: frequency, timing and impact on outcomes. EuroIntervention. 2012; 8(1): 62–70.

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By "Via Medica sp. z o.o." sp.k., ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: viamedica@viamedica.pl