Effects of pacemaker and implantable cardioverter defibrillator electrodes on tricuspid regurgitation and right sided heart functions

Peyman Arabi, Necla Özer, Ahmet Hakan Ateş, Hikmet Yorgun, Ali Oto, Kudret Aytemir
DOI: 10.5603/CJ.a2015.0060
·
Pubmed: 26412607
·
Cardiol J 2015;22(6):637-644.

open access

Vol 22, No 6 (2015)
Original articles
Submitted: 2015-05-17
Accepted: 2015-08-26
Published online: 2015-12-30

Abstract

Background: The aim of this study was to assess the effect of trans-tricuspid placement of permanent pacemaker (PPM), implantable cardioverter defibrillator (ICD) and cardiac resynchronization therapy (CRT) leads prospectively on tricuspid valve and right-sided heart functions using two-dimensional echocardiography.

Methods: A total of 41 patients (31 male, mean age: 63.6 ± 12.2 years) were included in this prospective study. Initial echocardiographic evaluation was performed before cardiac device implantation and re-evaluation by echocardiography was performed immediately after the procedure and at 1st, 6th and 12th months. In addition to standard echocardiographic examinations, vena contracta (VC), proximal isovelocity surface area (PISA), and tissue Doppler evaluations were also performed in the study population.

Results: Tricuspid regurgitation (TR) is worsened by 1 grade in 70.8% of the patients and 2 grades in 17.1% of the patients in the follow-up. Eight patients without baseline TR developed new-onset TR (9.8% mild, 9.8% moderate) after lead implantation. In the follow-up period, 41.5% of the patients who had mild TR before lead implantation developed moderate TR and 7.3% developed severe TR, whereas 19.5% of the patients with moderate TR developed severe TR during the follow-up. In the follow-up period, VC of TR was increased [median: 0.32 (0.16–0.60) cm in pre-implantation period, and 0.41 (0.18–0.80) cm at 12th month, p = 0.001]. Similarly PISA value of TR was also increased [median: 0.46 (0.15–1.10) cm in pre-implantation period and 0.52 (0.28–1.20) cm at 12th month, p = 0.001]. However, there is not a significant difference between PPMs/ICDs and CRTs regarding the effects on TR (p < 0.05). In addition, right ventricular dimensions and right atrial volumes were increased during the follow-up.

Conclusions: Implantation of permanent transvenous right ventricular electrode is associated with worsening of TR, right atrial and right ventricular dimensions. Further studies are needed in order to both outline the effect of those findings on outcomes and clarify the time dependent changes in those functions.

Abstract

Background: The aim of this study was to assess the effect of trans-tricuspid placement of permanent pacemaker (PPM), implantable cardioverter defibrillator (ICD) and cardiac resynchronization therapy (CRT) leads prospectively on tricuspid valve and right-sided heart functions using two-dimensional echocardiography.

Methods: A total of 41 patients (31 male, mean age: 63.6 ± 12.2 years) were included in this prospective study. Initial echocardiographic evaluation was performed before cardiac device implantation and re-evaluation by echocardiography was performed immediately after the procedure and at 1st, 6th and 12th months. In addition to standard echocardiographic examinations, vena contracta (VC), proximal isovelocity surface area (PISA), and tissue Doppler evaluations were also performed in the study population.

Results: Tricuspid regurgitation (TR) is worsened by 1 grade in 70.8% of the patients and 2 grades in 17.1% of the patients in the follow-up. Eight patients without baseline TR developed new-onset TR (9.8% mild, 9.8% moderate) after lead implantation. In the follow-up period, 41.5% of the patients who had mild TR before lead implantation developed moderate TR and 7.3% developed severe TR, whereas 19.5% of the patients with moderate TR developed severe TR during the follow-up. In the follow-up period, VC of TR was increased [median: 0.32 (0.16–0.60) cm in pre-implantation period, and 0.41 (0.18–0.80) cm at 12th month, p = 0.001]. Similarly PISA value of TR was also increased [median: 0.46 (0.15–1.10) cm in pre-implantation period and 0.52 (0.28–1.20) cm at 12th month, p = 0.001]. However, there is not a significant difference between PPMs/ICDs and CRTs regarding the effects on TR (p < 0.05). In addition, right ventricular dimensions and right atrial volumes were increased during the follow-up.

Conclusions: Implantation of permanent transvenous right ventricular electrode is associated with worsening of TR, right atrial and right ventricular dimensions. Further studies are needed in order to both outline the effect of those findings on outcomes and clarify the time dependent changes in those functions.

Get Citation

Keywords

tricuspid regurgitation, implantable cardiac devices, right heart functions, permanent pacemaker, implantable cardioverter defibrillator

About this article
Title

Effects of pacemaker and implantable cardioverter defibrillator electrodes on tricuspid regurgitation and right sided heart functions

Journal

Cardiology Journal

Issue

Vol 22, No 6 (2015)

Pages

637-644

Published online

2015-12-30

Page views

2097

Article views/downloads

2751

DOI

10.5603/CJ.a2015.0060

Pubmed

26412607

Bibliographic record

Cardiol J 2015;22(6):637-644.

Keywords

tricuspid regurgitation
implantable cardiac devices
right heart functions
permanent pacemaker
implantable cardioverter defibrillator

Authors

Peyman Arabi
Necla Özer
Ahmet Hakan Ateş
Hikmet Yorgun
Ali Oto
Kudret Aytemir

Regulations

Important: This website uses cookies. More >>

The cookies allow us to identify your computer and find out details about your last visit. They remembering whether you've visited the site before, so that you remain logged in - or to help us work out how many new website visitors we get each month. Most internet browsers accept cookies automatically, but you can change the settings of your browser to erase cookies or prevent automatic acceptance if you prefer.

By VM Media Group sp. z o.o., Grupa Via Medica, ul. Świętokrzyska 73, 80–180 Gdańsk, Poland
tel.:+48 58 320 94 94, fax:+48 58 320 94 60, e-mail: viamedica@viamedica.pl