The effectiveness of transvenous leads extractions implanted more than 10 years before
Abstract
Background: The increasing number of patients with cardiac implantable electronic devices (CIEDs) causes a rise in the absolute percentage of individuals qualifying for a transvenous lead extraction (TLE) due to infectious, vascular or lead failure related indications. As the survival time prolongs, TLE procedures more and more often concern the electrodes of long- -term functioning. Authors provide a retrospective analysis of the effectiveness and safety of TLE performed on leads implanted at least 10 years before the extraction.
Methods: Between 2008 and 2012 we performed TLE of 364 electrodes in 217 patients. Out of these, 66 (18.1%) leads in 43 (19.8%) patients had been implanted for at least 10 years. The mean dwelling time for electrodes was 161 months (120 to 330). In 62% of cases CIED-related infection was an indication for TLE. The following extracting techniques were used: manual direct traction, device traction, mechanical telescopic sheaths, autorotational cutting sheaths and femoral approach.
Results: Fifty-eight pacemakers and 8 defibrillating leads were extracted. Sixty-three (95%) completely, in the remaining 3 cases the clinical success was achieved with the small portion of the lead left into the vascular space. No major procedure complications were observed; minor complications were found in 3 (6%) patients.
Conclusions: TLE with the use of various endovascular techniques is an effective and safe method for treating infectious, vascular and mechanical complications of long-lasting CIEDs therapy.
Keywords: lead extractionpacing and defibrillation leads