Tissue Doppler echocardiography can be a useful technique to evaluate atrial conduction time
open access
Abstract
Background: The main purpose of this study is to determine the correlation of inter- and intraatrial conduction times between the electrophysiological and tissue Doppler echocardiographic measurements, and to evaluate the appropriateness of tissue Doppler echocardiography for this measurement.
Methods: One-hundred and one patients were included in the study who underwent electrophysiological study for clinical arrhythmias. Inter- and intraatrial conduction times were measured from intracardiac electrograms. Atrial conduction times were also measured by tissue Doppler echocardiography by evaluating atrial electromechanical delay between lateral mitral annulus, septal mitral annulus, and right ventricular tricuspid annulus. The correlation between electrophysiological and echocardiographic atrial conduction times were analyzed.
Results: We found a weak correlation between the measurements of interatrial conduction times with the electrophysiological and tissue Doppler techniques (r = 0.308; p = 0.002). The correlation for intraleft atrial conduction times was moderate (r = 0.652; p < 0.001). There was no correlation between the measurements of intra-right atrial conduction times.
Conclusions: We concluded that tissue Doppler echocardiography can be used for the measurement of interatrial and intra-left atrial conduction times. Tissue Doppler echocardiography can be a suitable technique to evaluate atrial substrate. (Cardiol J 2012; 19, 5: 487-493)
Abstract
Background: The main purpose of this study is to determine the correlation of inter- and intraatrial conduction times between the electrophysiological and tissue Doppler echocardiographic measurements, and to evaluate the appropriateness of tissue Doppler echocardiography for this measurement.
Methods: One-hundred and one patients were included in the study who underwent electrophysiological study for clinical arrhythmias. Inter- and intraatrial conduction times were measured from intracardiac electrograms. Atrial conduction times were also measured by tissue Doppler echocardiography by evaluating atrial electromechanical delay between lateral mitral annulus, septal mitral annulus, and right ventricular tricuspid annulus. The correlation between electrophysiological and echocardiographic atrial conduction times were analyzed.
Results: We found a weak correlation between the measurements of interatrial conduction times with the electrophysiological and tissue Doppler techniques (r = 0.308; p = 0.002). The correlation for intraleft atrial conduction times was moderate (r = 0.652; p < 0.001). There was no correlation between the measurements of intra-right atrial conduction times.
Conclusions: We concluded that tissue Doppler echocardiography can be used for the measurement of interatrial and intra-left atrial conduction times. Tissue Doppler echocardiography can be a suitable technique to evaluate atrial substrate. (Cardiol J 2012; 19, 5: 487-493)
Keywords
atrial conduction time; tissue Doppler echocardiography


Title
Tissue Doppler echocardiography can be a useful technique to evaluate atrial conduction time
Journal
Issue
Pages
487-493
Published online
2012-10-06
Page views
1587
Article views/downloads
1954
DOI
10.5603/CJ.2012.0089
Bibliographic record
Cardiol J 2012;19(5):487-493.
Keywords
atrial conduction time
tissue Doppler echocardiography
Authors
Ali Deniz
Levent Sahiner
Kudret Aytemir
Baris Kaya
Giray Kabakci
Lale Tokgozoglu
Ali Oto